Impact of detector selections on inter‐institutional variability of flattening filter‐free beam data for TrueBeam? linear accelerators

机译:探测器选择对Truebeam的扁平滤波器梁数据的机构间变异性的影响吗?线性加速器

获取原文
   

获取外文期刊封面封底 >>

       

摘要

This study evaluates the type of detector influencing the inter‐institutional variability in flattening filter‐free (FFF) beam‐specific parameters for TrueBeam? linear accelerators (Varian Medical Systems,Palo Alto, CA, USA). Twenty‐four beam data sets, including the percent depth dose (PDD), off‐center ratio (OCR), and output factor (OPF) for modeling within the Eclipse (Varian Medical Systems) treatment planning system, were collected from 19 institutions. Although many institutions collected the data using CC13 (IBA Dosimetry, Schwarzenbruck, Germany) or PTW31010 semiflex (PTW Freiburg, Freiburg, Germany) ionization chambers, some institutions used diode detectors, diamond detectors, and ionization chambers with smaller cavities. The OCR data included penumbra width, full width at half maximum (FWHM), and FFF beam‐specific parameters, including unflatness and slope. The data measured by CC13/PTW31010 ionization chambers were compared with those measured by all other detectors. PDD data demonstrated the variations within ±1% at the dose fall‐off region deeper than peak depth. The penumbra widths of the OCR measured with the CC13/PTW31010 detectors were significantly larger than those measured with all other detectors ( P ?0.05). Especially the EDGE detector (Sun Nuclear Corp., Melbourne, FL, USA) and the microDiamond detectors (model 60019; PTW Freiburg) demonstrated much smaller penumbra values compared to those of the CC13/PTW31010 detectors for the 30?×?30?mmsup2/sup field. There was no difference in the FWHM, unflatness, and slope parameters between the values for the CC13/PTW31010 detectors and all other detectors. OPF curves demonstrated small variations, and the relative difference from the mean value of each data point was almost within 1% for all field sizes. Although the penumbra region exhibited detector‐dependent variations, all other parameters showed tiny interunit variations regardless of the detector type.
机译:本研究评估影响无滤网(FFF)光束特定参数的机构间变异性的探测器的类型?线性加速器(Varian医疗系统,帕洛阿尔托,加州,美国)。从19个机构收集了二十四个光束数据集,包括用于在Eclipse(Varian Medical Systems)治疗系统内进行建模的深度剂量(PDD),偏心比(OCR)和输出因子(OPF)。虽然许多机构使用CC13(IBA Dosimetry,Schwarzenbruck,Germany)或PTW31010 Semiflex(PTW Freiburg,Freiburg,德国)电离室,一些机构使用二极管探测器,金刚石探测器和具有较小腔的电离室的数据。 OCR数据包括Penumbra宽度,半宽度为半最大(FWHM)和FFF光束的参数,包括unflat度和斜率。将CC13 / PTW31010电离室测量的数据与所有其他探测器测量的数据进行了比较。 PDD数据在比峰值深度更深的剂量下降区域的±1%内显示±1%的变化。使用CC13 / PTW31010检测器测量的OCR的PENUMBRA宽度明显大于用所有其他探测器测量的探测器(P <0.05)。特别是边缘探测器(Sun核核公司,墨尔本,FL,USA)和Microdiamond探测器(型号60019; PTW Freiburg)与CC13 / PTW31010探测器的探测器相比展示了更小的半影值,用于30?×30?mm 2 字段。 CC13 / PTW31010检测器和所有其他检测器的值之间的FWHM,unllat度和斜率参数没有差异。 OPF曲线显示出小的变化,并且每个数据点的平均值的相对差异几乎在1%范围内为所有字段大小。虽然PENUMBRA区域表现出依赖于探测器依赖性变化,但所有其他参数都显示出微小的间歇变化,而不管探测器类型。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号