Characteristic Features of an Analytical Column with a PentafluorophenylpropylStationary Phase Applied To a Determination of a Fluorinated Phenyl Alanyl Derivative Compound, Gw823093, In Human Urine Usingan Lc-Esi-Ms/Ms Method

机译:具有五氟苯丙基己酰基的分析塔的特征,应用于氟化苯丙二醇衍生物,GW823093,在人尿中使用的LC-ESI-MS / MS法测定

获取原文

摘要

Pentafluorophenylpropyl (PFPP) stationary phases allow for dipole-dipole, pi-pi, charge transfer and ion- exchange interactions and have shown unique retention of small, polar analytes. PFPP columns are able to retain protonated bases in the presence of high percentages of an organic modifier, which is a suitable condition for liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS) analysis. Taking advantage of the characteristics of PFPP columns, a liquid chromatography-electrospray ionization coupled with tandem mass spectrometry (LC-ESI-MS/MS) method was developed to quantify the concentration of a fluorinated phenylalanyl derivative compound, GW823093, in human urine. The analyte is retained with great difficulty on an octadecylsilane- bonded silica (ODS) column, but was strongly retained on a PFPP column in 5 mmol/L ammonium formate buffer (pH 2.9) even with an acetonitrile concentration as high as 80%; these conditions protonated the analyte and enabled higher MS signal intensity. However, the analyte peak shape deteriorated after repeated sample injections when urine samples were analyzed under the selected conditions. To solve this problem, an on-line sample clean-up process with an ODS column was employed before the sample was introduced to the PFPP analytical column. A good peak shape without ion suppression was achieved when using acidified buffer containing 10% acetonitrile as a washing solvent for the process, suggesting that components retained by ionic and weakly hydrophobic interactions were removed. This method was optimized and was linear in quantitation in the range of 25-5000 ng/mL, with a mean coefficient of determination over 0.9998. The validated method was successfully applied to a clinical study of GW823093.
机译:五氟苯基丙基(PFPP)固定相可以允许偶极偶极子,PI-PI,电荷转移和离子交换相互作用,并显示出小极性分析物的独特保留。 PFPP柱能够在高百分比的有机改性剂的存在下保持质子化碱基,这是液相色谱与串联质谱(LC-MS / MS)分析的合适条件。利用PFPP柱的特性,开发了与串联质谱(LC-ESI-MS / MS)偶联的液相色谱 - 电喷雾电离,以量化氟化苯丙烯基衍生物化合物,GW823093,人尿中的浓度。在十八烷基硅烷粘合的二氧化硅(ODS)柱上,分析物非常困难,但是在5mmol / L铵甲酸铵缓冲液(pH2.9)中强烈保留在PFPP柱上,即使乙腈浓度高达80%;这些条件质量对分析物质量并使能更高的MS信号强度。然而,当在所选条件下分析尿液样品时,在重复的样品注射后劣化的分析物峰值。为了解决这个问题,在将样品引入PFPP分析柱之前使用具有ODS柱的在线样品清理方法。当使用含有10%乙腈的酸化缓冲液作为该方法的洗涤溶剂时,达到了没有离子抑制的良好峰形状,表明通过离子保留的成分和弱疏水相互作用。该方法进行了优化,在定量中线性在25-5000ng / ml的范围内,均为0.9998的平均测定系数。经过验证的方法已成功应用于GW823093的临床研究。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号