In vivo performance of near-infrared light transillumination for dentine proximal caries detection in permanent teeth

机译:在体内

获取原文
   

获取外文期刊封面封底 >>

       

摘要

ObjectiveThe purpose of thisin vivostudy was to confirm the detection of proximal caries using near-infrared light transillumination (NILTI) (DIAGNOcam) device, and to compare the diagnostic performance of the device with other caries detection methods, including visual examination using the International Caries Detection and Assessment System (ICDAS), bitewing radiography (BW), an LED-based device (Midwest Caries I.D.), and a laser fluorescence device (LFpen).MethodsA total of 974 proximal surface of permanent posterior teeth from 34 patients (19 females and 15 males between the ages of 22–55) were evaluated in the present study. After clinical examination of each proximal surface by an experienced examiner, they were coded according to the ICDAS criteria and subsequently with BW, the NILTI, LED, and LFpen. The proximal dentin caries of 106 were opened based on the combination of visual, NILTI and radiographic assessment and validated; and were treated with restorative materials. The data were analyzed with descriptive statistics and calculation of the sensitivity, specificity, and area under curve (AUC).ResultsThe highest sensitivity values were recorded from NILTI readings (99.1%), followed by BW scores (86.8%). The highest specificity values were recorded from ICDAS (100%). The best AUC values were found from the NILTI readings (0.97), followed by BW (0.93) and ICDAS (0.87).ConclusionThe NILTI device exhibited the best performance in terms of detecting proximal dentin caries.
机译:该目的的目的是使用近红外光透射度(NILTI)(NILTI)(诊断电池)装置来确认近端龋齿的检测,并比较器件与其他龋齿检测方法的诊断性能,包括使用国际龋齿检测的视觉检查和评估系统(ICDAS),BITEWING放射线照相(BW),基于LED的设备(中西部龋ID)和激光荧光装置(LFPEN).Methodsa总共34名患者的永久性后齿的974个近端表面(19名女性和在本研究中评估了22-55岁之间的15岁的男性。经验丰富的审查员临床检查每个近端表面后,它们根据ICDAS标准编码,随后与BW,NILTI,LED和LFPen进行编码。基于视觉,Nilti和射线照相评估和验证的组合,打开106的近端牙本质龋;并用恢复材料处理。通过描述性统计分析数据,并计算曲线下的敏感性,特异性和面积的计算,从Nilti读数(99.1%)记录了最高灵敏度值,其次是BW分数(86.8%)。从ICDA(100%)记录最高的特异性值。从Nilti读数(0.97)中发现了最好的AUC值,然后是BW(0.93)和ICDA(0.87)。结论NILTI器件在检测近端牙本质龋齿方面表现出最佳性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号