Nexfin Noninvasive Continuous Hemodynamic Monitoring: Validation against Continuous Pulse Contour and Intermittent Transpulmonary Thermodilution Derived Cardiac Output in Critically Ill Patients

机译:Nexfin非侵入性连续血液动力学监测:对连续脉冲轮廓的验证和间歇性的经稳态患者的衍生心输出

获取原文
           

摘要

Introduction. Nexfin (Bmeye, Amsterdam, Netherlands) is a noninvasive cardiac output (CO) monitor based on finger arterial pulse contour analysis. The aim of this study was to validate Nexfin CO (NexCO) against thermodilution (TDCO) and pulse contour CO (CCO) by PiCCO (Pulsion Medical Systems, Munich, Germany).Patients and Methods. In a mix of critically ill patients (n=45), NexCO and CCO were measured continuously and recorded at 2-hour intervals during the 8-hour study period. TDCO was measured at 0–4–8 hrs.Results. NexCO showed a moderate to good (significant) correlation with TDCO (R20.68,P<0.001) and CCO (R20.71,P<0.001). Bland and Altman analysis comparing NexCO with TDCO revealed a bias (± limits of agreement, LA) of 0.4 ± 2.32 L/min (with 36% error) while analysis comparing NexCO with CCO showed a bias (±LA) of 0.2 ± 2.32 L/min (37% error). NexCO is able to follow changes in TDCO and CCO during the same time interval (level of concordance 89.3% and 81%). Finally, polar plot analysis showed that trending capabilities were acceptable when changes in NexCO (ΔNexCO) were compared toΔTDCO andΔCCO (resp., 89% and 88.9% of changes were within the level of 10% limits of agreement).Conclusion. we found a moderate to good correlation between CO measurements obtained with Nexfin and PiCCO.
机译:介绍。 Nexfin(Bmeye,Amsterdam,荷兰)是一种基于手指动脉脉冲轮廓分析的非侵入性心输出(CO)监测。本研究的目的是通过PICCO(武器医疗系统,慕尼黑,德国)。验证NEXFIN CO(NEXCO)对热稀土(TDCO)和脉冲轮廓CO(CCO).Patient和方法。在危重患者(n = 45)的混合物中,在8小时的研究期间连续测量Nexco和CCO并以2小时的间隔记录。 TDCO以0-4-8小时测量。结果。 Nexco与TDCO(R20.68,P <0.001)和CCO(R20.71,P <0.001)显示了中等至良好(显着的)相关性。 Bland和Altman分析将Nexco与TDCO进行比较,显示出0.4±2.32 L / min(具有36%误差)的偏差(协议,LA),同时分析NEXCO与CCO的偏置(±LA)为0.2±2.32升/ min(37%错误)。 Nexco能够在同一时间间隔内遵循TDCO和CCO的变化(一定程度的级别89.3%和81%)。最后,极性绘图分析表明,当与Δtdco和ΔCCO(ΔChco和ΔCCO(RESP)进行比较时,当NEXCO(Δnexco)的变化(ΔCHCO)的变化(REB值为89%和88.9%的变化程度在10%的协议范围内)。结论。我们发现用Nexfin和Picco获得的CO测量与CO测量之间的良好相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号