首页> 外文期刊>Respiratory Research >Performance of a new hand-held device for exhaled nitric oxide measurement in adults and children
【24h】

Performance of a new hand-held device for exhaled nitric oxide measurement in adults and children

机译:在成人和儿童中呼出一氧化氮测量的新手持装置的性能

获取原文
       

摘要

BackgroundExhaled nitric oxide (NO) measurement has been shown to be a valuable tool in the management of patients with asthma. Up to now, most measurements have been done with stationary, chemiluminescence-based NO analysers, which are not suitable for the primary health care setting. A hand-held NO analyser which simplifies the measurement would be of value both in specialized and primary health care. In this study, the performance of a new electrochemical hand-held device for exhaled NO measurements (NIOX MINO) was compared with a standard stationary chemiluminescence unit (NIOX).MethodsA total of 71 subjects (6–60 years; 36 males), both healthy controls and atopic patients with and without asthma were included. The mean of three approved exhalations (50 ml/s) in each device, and the first approved measurement in the hand-held device, were compared with regard to NO readings (Bland-Altman plots), measurement feasibility (success rate with 6 attempts) and repeatability (intrasubject SD).ResultsSuccess rate was high (≥ 84%) in both devices for both adults and children. The subjects represented a FENO range of 8–147 parts per billion (ppb). When comparing the mean of three measurements (n = 61), the median of the intrasubject difference in exhaled NO for the two devices was -1.2 ppb; thus generally the hand-held device gave slightly higher readings. The Bland-Altman plot shows that the 95% limits of agreement were -9.8 and 8.0 ppb. The intrasubject median difference between the NIOX and the first approved measurement in the NIOX MINO was -2.0 ppb, and limits of agreement were -13.2 and 10.2 ppb. The median repeatability for NIOX and NIOX MINO were 1.1 and 1.2 ppb, respectively.ConclusionThe hand-held device (NIOX MINO) and the stationary system (NIOX) are in clinically acceptable agreement both when the mean of three measurements and the first approved measurement (NIOX MINO) is used. The hand-held device shows good repeatability, and it can be used successfully on adults and most children. The new hand-held device will enable the introduction of exhaled NO measurements into the primary health care.
机译:背景技术已经显示出氧化氮(NO)测量是在哮喘患者管理中的有价值的工具。到目前为止,大多数测量都是用静止的化学发光的没有分析仪进行的,这不适合初级保健环境。手持式没有分析仪,简化了测量的分析仪在专业和初级保健中都具有价值。在这项研究中,将用于呼出的新型电化学手持装置的性能与标准固定化学发光单元(NIOX)进行比较.Methodsa总共71个受试者(6-60岁; 36个男性)包括健康对照和有没有哮喘的特应性患者。在每个装置中的三种经批准的呼气(50mL / s)和手持装置中的第一次批准的测量值相同,在没有读数(Bland-Altman Plots),测量可行性(成功率,6次尝试)和成人和儿童的两种设备中,重复性(intrAsbject SD).Resultsuccescess率高(≥84%)。受试者代表FENO范围为8-147分(PPB)。比较三次测量的平均值(n = 61)时,两种器件呼出的呼出差差的中位数为-1.2ppb;因此,通常,手持装置略高读取。 Bland-Altman Plot表明,95%的协议限额为-9.8和8.0 ppb。 NIOX和NIOX MINO中的第一批准测量之间的胃胆机中值差异为-2.0ppb,协议限制为-13.2和10.2 ppb。 NIOX和NIOx MINO的中值重复性分别为1.1和1.2 PPB。结论手持装置(NIOX MINO)和静止系统(NIOX)在临床上可接受的一致,当三次测量和第一批准的测量(使用了niox mino。手持设备显示出良好的重复性,可在成人和大多数儿童上成功使用。新的手持设备将使呼出的没有测量的引入初级保健。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号