An Analysis Method of Dose Error Caused by the Couch Tilt of CT Simulation Machine

机译:CT模拟机沙发倾斜引起的剂量误差分析方法

获取原文
           

摘要

Purpose: If the horizontal plane is inconsistent between computer tomography (CT) couch and line accelerator (LA) couch, the position of the isocenter point in the radiotherapy treatment planning system (TPS) should deviate from that of it on the accelerator couch board. Then, the actual dose distribution was different from the radiotherapy plan we designed in TPS. This paper introduces a method to evaluate the dose deviation caused by the tilt of CT couch. Methods: We calculated the isocenter points of the position coordinates both in CT couch and LA couch, and compared the dose distribution when the twopositions of isocenter points were applied in radiotherapy plan independently in TPS. The dose distribution difference of a breast radiotherapyplan was analyzed as a demonstration with this method. Results: The distance between the two isocenter positions increased with angle of CT couch. The tilt of the couch had an impact on the dose distribution, especially in larynx's maximum dose parameter. Conclusion: This method could quantitatively analyze the dose distribution deviationcaused by the tilt of CT couch plate. The results can provide a valuable suggestion for clinical medical strategy.
机译:目的:如果电脑断层扫描(CT)沙发和线路加速器(LA)沙发之间的水平平面不一致,则在放射疗法治疗计划系统(TPS)中的Isocenter点的位置应偏离加速器沙发板上的位置。然后,实际剂量分布与我们在TPS中设计的放射疗法计划不同。本文介绍了一种评价CT沙发倾斜引起的剂量偏差的方法。方法:我们计算了CT沙发和La沙发中的位置坐标的等离心点,并将剂量分布与在TPS中独立于放射疗法计划中应用。用这种方法分析乳房放射疗法的剂量分布差异作为证明。结果:两个等中心位置之间的距离随着CT沙发的角度而增加。沙发的倾斜对剂量分布产生影响,特别是在喉的最大剂量参数中。结论:该方法可以定量分析CT沙发板倾斜偏离的剂量分布。结果可以为临床医疗策略提供有价值的建议。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号