首页> 外文期刊>Neurospine. >A New Shielding Curtain for Protection of Intraoperative Radiation During Minimally Invasive Spine Surgery
【24h】

A New Shielding Curtain for Protection of Intraoperative Radiation During Minimally Invasive Spine Surgery

机译:一种新的屏蔽窗帘,用于保护术中辐射在微创脊柱手术期间

获取原文
           

摘要

Objective:Gradually increasing number of minimally invasive spine surgery (MISS), there is an increasing risk of radiation exposure to medical personnel during the surgery. We measured the radiation exposure of the operating room personnel during MISS, tried to find the riskiest person, and checked the effectiveness of a new lead-composite shielding curtain.Methods:Radiation exposure of medical staffs (operator, first assistant, anesthesiologist, and scrub nurse) involved in MISS procedures of 35 patients without shielding curtain (nonshield group) and 35 patients with shielding curtain (shield group). The shielding curtain had 0.25-mm nominal lead equivalent and was mounted on 2 frame bars fixed on the operating table.Results:In the nonshield group, radiation exposure was significantly higher in the order of operator first assistant scrub nurse anesthesiologist (p 0.001) during both anteroposterior (AP) and lateral views. In the shield group, the radiation exposure of the operator and the scrub nurse decreased significantly by 94.1% and 76.4% in AP view (p 0.001), and by 96.3% and 73.9% in lateral view (p 0.001), respectively.Conclusion:Since the radiation dose of the operator was highest in a C-arm-guided MISS, there is a high priority need to protect the operator from the radiation exposure. The shielding curtain could most effectively reduce the radiation exposure of the operator.
机译:目的:逐步越来越多的侵袭性脊柱手术(Miss),在手术期间辐射暴露的风险越来越大。我们在小姐期间测量了手术室人员的辐射曝光,试图找到最风险的人,并检查了新的铅综合屏幕窗帘的有效性。方法:医务人员的辐射曝光(操作员,第一助手,麻醉师和磨砂护士)参与35名患者的小姐程序,无需屏蔽窗帘(非屏蔽组)和35名屏蔽窗帘(盾牌组)。屏蔽窗帘具有0.25毫米的标称铅等同物,并安装在操作表上的2个框架杆上。结果:在非屏蔽组中,辐射曝光按操作员的顺序显着提高>第一个助手>磨砂护士>麻醉学者(P <0.001)在前后剂(AP)和侧面视图中。在屏蔽组中,透射液和擦洗护士的辐射暴露在AP视图(P <0.001)中显着降低了94.1%和76.4%,分别在横向视图(P <0.001)中达到96.3%和73.9%。结论:由于操作员的辐射剂量在C形臂导向错过中最高,因此有一个高优先级需要保护操作员免受辐射暴露。屏蔽窗帘最有效地减少操作员的辐射曝光。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号