This paper focuses on the problematic of intraocular pressure (IOP) measurements, performed by non-invasive methods. More specifically, the devices that are connected with the presented finding are non-contact tonometers that use concentrated air stream and optical sensors to determine the IOP within a human’s eye. The paper analyzes various influential factors that have an effect on the determination of the IOP values originating from the patients themselves and from the non-contact tonometer devices. The paper furthermore elaborates on the lack of independent methods of calibration and control of these devices. In order to fill this gap a measurement standard device that is capable of calibrating and testing these devices with traceability to the basic SI unit is presented. A detailed characterization and the determination of the expected uncertainty of the device are provided. By introducing an independent and traceable calibration method and control of non-contact tonometers into the clinical practice, the reliability of the measured IOP that is the primary indicator of glaucoma can be improved.
机译:本文侧重于通过非侵入性方法进行的眼压(IOP)测量的问题。更具体地,与所提出的发现连接的设备是使用集中的空气流和光学传感器的非接触式调谐,以确定人眼内的IOP。本文分析了各种影响因素,这些因素对源自患者本身的IOP值和非接触式调谐计器件的影响有影响。本文还详细阐述了对这些装置的校准和控制缺乏独立的方法。为了填充该间隙,提供了能够通过可追溯性校准和测试这些设备的测量标准设备,该装置具有可追溯性对基本SI单元。提供了详细的表征和确定装置的预期不确定性。通过引入独立且可追踪的校准方法和对临床实践的非接触式校准方法和对临床实践的可靠性,可以提高是青光眼的主要指示剂。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号