首页> 外文期刊>Medicine. >Does nonexistent of your hands on the screen guarantee no radiation exposure to your body? – Study on exposure of the practitioner's hands to radiation during C-arm fluoroscopy-guided injections and effectiveness of a new shielding device
【24h】

Does nonexistent of your hands on the screen guarantee no radiation exposure to your body? – Study on exposure of the practitioner's hands to radiation during C-arm fluoroscopy-guided injections and effectiveness of a new shielding device

机译:屏幕上的手是否不存在,保证无辐射暴露在您的身体上? - 在新屏蔽装置的C形臂透视引导注射过程中,从业者手辐射的辐射研究

获取原文
       

摘要

Observational phantom study . This study aimed to evaluate the radiation exposure dose of practitioner 's hands when performing C-arm guided procedures and to determine the usefulness of our newly designed radiation shielding device . C-arm guided procedures including lumbar transforaminal epidural steroid injections (TFESIs) are commonly used for pain control induced by lumbar radiculopathy. The practitioner 's hands are vulnerable to radiation exposure because of the long exposure time and short distance from the radiation resource. No studies to date have reported the cumulative exposure of the physician's hands according to location and exposure time. Using a chest phantom irradiated with X-rays under lumbar TFESI conditions, cumulative scatter radiation dose was measured at 36 points using a dosimeter. The measurements were checked at 1, 3, 5, 10 minutes of radiation exposure . The experiment was repeated using our newly designed shielding device . Significant radiation accumulation was observed in the field where the practitioner 's hands might be placed during C-arm guided procedures. The further the distance from the radiation resource and the shorter the exposure time, the smaller was the cumulative radiation expose dose. The new shielding device showed an excellent shielding rate (66.0%–99.9%) when the dosimeter was within the shielding range. However, at some points, increased accumulated radiation exposure dose was observed, although the dosimeter was within the range of the shielding device . To reduce radiation exposure of the practitioner 's hands when performing C-arm -guided procedures, the radiation exposure time should be decreased and a greater distance from the radiation resource should be maintained. When using our shielding device , placing the hand close to the device surface and minimizing the time using fluoroscopy minimized the radiation exposure of the hand.
机译:观察幻影研究。该研究旨在在进行C形臂导向程序时评估从业者手的辐射曝光剂量,并确定我们新设计的辐射屏蔽装置的用途。 C-ARM引导程序包括腰椎突变体硬膜外类固醇注射(TFERSIES)通常用于腰部诱导的疼痛对照。由于距离辐射资源的曝光时间很长,从业者的手容易受到辐射曝光。迄今为止没有研究报告了医师手的累积暴露,根据位置和暴露时间。使用腰部TFESI条件下用X射线照射的胸部幻影,使用剂量计在36点测量累积散射辐射剂量。在1,3,5,10分钟的辐射暴露中检查测量。使用新设计的屏蔽装置重复实验。在C形臂导向程序期间,在该领域中观察到显着的辐射积累。累积辐射越小的辐射资源和较短曝光时间的距离进一步较小的距离。当剂量计在屏蔽范围内时,新的屏蔽装置显示出优异的屏蔽速率(66.0%-99.9%)。然而,在某些点处,观察到增加的累积辐射曝光剂量,尽管剂量计在屏蔽装置的范围内。为了减少从业者手中的辐射曝光,在执行C-ARM-umid程序时,应减少辐射曝光时间,应保持距离辐射资源的距离。使用我们的屏蔽装置时,将手靠近器件表面并最小化使用荧光检查的时间最小化手的辐射曝光。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号