首页> 外文期刊>International Journal of Dentistry >Laboratory Rational of Changes in the Crystal Lattice of Nickel-Titanium Endodontic Rotary Files in Autoclaving
【24h】

Laboratory Rational of Changes in the Crystal Lattice of Nickel-Titanium Endodontic Rotary Files in Autoclaving

机译:高压灭菌中镍钛牙髓旋转文件晶格晶格的改变实验室

获取原文
获取外文期刊封面目录资料

摘要

Objectives. The aim of this study was to investigate the changes of the crystal lattice of NiTi instruments after repeated autoclave sterilization cycles based on the results to conclude about the influence of multiple sterilization on the characteristics of ProTaper clinical use. Methods. 21 samples of ProTaper Universal rotary files were divided into 3 groups of 7 samples. After 1, 4, and 7 cycles of sterilization, the samples were observed using еnergy-dispersive X-ray spectroscopy (EDX), X-ray powder diffraction (XRD), and scanning electron microscopy (SEM). Results. EDS analysis confirmed that all samples were composed mainly of nickel and titanium, also Fe (presumably steel), Cr, Co, and Zn were found. The percentage of nickel and titanium was affected by repeated sterilization cycles. The mass fraction of Fe, Co, Zn, and Cr decreased after 1, 4, and 7 sterilization cycles. According to the results of the second study, it was found that, with increasing autoclaving cycles, the percentage of Fe decreased. There were changes in the three obtained diffractograms, which indicated an increase in the number of the martensitic phase and a decrease in austenitic. The cutting efficiency of the ProTaper is also reduced during repeated sterilization cycles, which can be caused by the austenitic and martensitic phase displacement. Conclusion. Manufacturing features and repeated sterilization cycles increase the internal deformation in the structure of NiTi alloy, which increases the risk of fracture.
机译:目标。本研究的目的是研究重复高压灭菌循环后NITI仪器晶格的变化,结果是关于多次灭菌对蛋白临床用途的影响。方法。将21个蛋白通用旋转文件样本分为3组7个样品。在1,4和7个杀菌循环之后,使用X射线X射线光谱(EDX),X射线粉末衍射(XRD)和扫描电子显微镜(SEM)观察样品。结果。 EDS分析证实,所有样品主要由镍和钛组成,也发现Fe(大规模的钢),Cr,CO和Zn。镍和钛的百分比受重复灭菌循环的影响。在1,4和7个灭菌循环后,Fe,Co,Zn和Cr的质量分数降低。根据第二种研究的结果,发现随着高压灭菌循环的增加,Fe的百分比减少。三个获得的衍射图中存在变化,表明马氏体相的数量增加和奥氏体减少。在重复的灭菌循环期间,蛋白的切割效率也降低,这可能是由奥氏体和马氏体相位置换引起的。结论。制造特征和重复的灭菌循环增加了Niti合金结构的内部变形,这增加了骨折的风险。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号