首页> 外文期刊>Allergy, Asthma & Clinical Immunology >Hooked epinephrine auto-injector devices in children: four case reports with three different proposed mechanisms
【24h】

Hooked epinephrine auto-injector devices in children: four case reports with three different proposed mechanisms

机译:儿童钩住的肾上腺自动注射器装置:四个案例报告,具有三种不同的提出机制

获取原文
           

摘要

The prevalence of epinephrine auto-injectors (EAI) use is on the rise. Our objective was to describes children with hooked EAI needles that were embedded in soft tissues. Results: Two children self-injected in their shins. The embedded EAIs required removal in the Emergency Department. Both needles were hooked and splayed at the tip. A boy in anaphylaxis kicked his leg during EAI injection and the hooked needle embedded under his skin and was difficult to dislodge. The exposed needle was curved. A girl had an EAI administered for anaphylaxis, which was also difficult to dislodge. On removal, the distal needle tip was hooked approximately 160 degrees. Images of the device revealed that the needle fired off-center from the device and the device components were cracked. We propose three different explanations for these hooked EAI needles. The first is that the needle could hit bone during injection and curve rather than penetrates further. Secondly, the needle could bend when the patient moves during injection. Thirdly, if a needle fires sufficiently off-center to hit the cartridge carrier, this could hook the needle prior to injection. Awareness of the reasons for needle hooking, damage observed, and challenges and successful approaches to their removal, can better prepare the provider for these uncommon events. Teaching parents, children and educators about safe EAI storage and appropriate restraint during use may prevent some of these accidental injuries. Reporting device failures may lead to improvements in device performance and design.
机译:肾上腺素自动注射器(EAI)使用的患病率在上升。我们的目标是将迷住于软组织中的EAI针头的儿童描述。结果:两名儿童在他们的胫骨上自我注射。嵌入式的EAIS需要在急诊部门搬迁。两只针被钩住并在尖端张开。在EAI注射期间,一个男孩在EAI注射时踢了他的腿,并且嵌入着皮肤下的钩针却难以脱落。暴露的针是弯曲的。一个女孩对过敏反应的EAI,这也难以脱落。去除后,远端针尖约160度。设备的图像显示,从装置中射出射击中心,并且将器件部件破裂。我们为这些钩住的EAI针提出了三种不同的解释。首先是针可以在注射和曲线期间击中骨,而不是进一步渗透。其次,当患者在注射期间移动时,针可以弯曲。第三,如果针头充分偏离中心以击中盒式磁带,这可以在注射之前钩钩。意识到针钩钩的原因,观察到的伤害以及挑战和成功的拆除方法,可以更好地为这些罕见的事件做好准备。教授父母,儿童和教育工作者关于安全EAI储存和使用期间适当的克制可能会阻止其中一些意外伤害。报告设备故障可能导致设备性能和设计的改进。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号