The sensitivity of different methods for detecting abnormalities in auditory nerve function

机译:不同方法检测听觉神经函数异常的敏感性

获取原文
           

摘要

Cochlear implants (CIs) have become important for the treatment of severe-to-profound sensorineural hearing loss (SNHL). Meanwhile, electrically evoked compound action potentials (ECAPs) and electrically evoked auditory brainstem responses (EABRs), which can be examined and evaluated with minimal patient cooperation, have become more reliable for tone measurement and speech recognition postoperatively. However, few studies have compared the electrophysiological characteristics of the auditory nerve using ECAPs and EABRs under different functional states of the auditory nerve (FSANs). We used guinea pig models in which six electrodes were implanted unilaterally with continuous electrical stimulation (ES) for 4?h. The amplitude growth functions (AGFs) of the alternating polarity ECAP (AP-ECAP) and forward-masking subtraction ECAP (FM-ECAP), as well as the EABR waves under “normal” and “abnormal” FSANs, were obtained. Both the AP-ECAP and FM-ECAP thresholds were significantly higher than those measured by EABR under both “normal” FSAN and “abnormal” FSANs (p??0.05). There was a significant difference in the slope values between electrodes 1 and 2 and electrodes 3 and 4 in terms of the AP-ECAP under the “abnormal” FSAN (p??0.05). The threshold gaps between the AP-ECAP and FM-ECAP were significantly larger under the “abnormal” FSAN than under the “normal” FSAN (p??0.05). Both of the ECAP thresholds were higher than the EABR thresholds. The AP-ECAP was more sensitive than the FM-ECAP under the “abnormal” FSAN.
机译:耳蜗植入物(CIS)对治疗严重致敬的感觉神经听力损失(SNHL)变得重要。同时,电诱发的复合作用电位(ECAPS)和电诱发听觉脑干响应(EABRS)可以通过最小的耐心合作进行检查和评估,对音调测量和术后的语音识别变得更加可靠。然而,很少有研究在听觉神经(FSANs)的不同功能状态下使用ECAPS和EABRS比较了听觉神经的电生理学特征。我们使用了几内亚猪模型,其中六个电极单侧植入连续电刺激4℃。获得了交替极性ECAP(AP-ECAP)和前掩蔽减法ECAP(FM-ECAP)的幅度生长功能(AGF)以及“正常”和“异常”FSAN下的EABR波。 AP-ECAP和FM-ECAP阈值均明显高于EABR在“正常”FSAN和“异常”FSAN下测量的阈值(P?<?0.05)。在“异常”FSAN下的AP-ECAP方面,电极1和2和电极3和4之间存在显着差异,在“异常”FSAN下(P?<β05)。在“异常”FSAN下,AP-ECAP和FM-ECAP之间的阈值间隙明显更大,而不是“正常”FSAN(P?<?0.05)。 eCAP阈值都高于EABR阈值。 AP-ECAP比“异常”FSAN下的FM-ECAP更敏感。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号