Ventricular late potentials are low-amplitude signals originating from damaged myocardium and detected on the body surface by ECG filtering and averaging. Digital filters present in commercial equipment may interfere with the ability of arrhythmia stratification. We compared 40-Hz BiSpec (BI) and classical 40- to 250-Hz band-pass Butterworth bidirectional (BD) filters in terms of impact on time domain variables and diagnostic properties. In a transverse retrospective age-adjusted case-control study, 221 subjects with sinus rhythm without bundle branch block were divided into three groups after signal-averaged ECG acquisition: GI (N = 40), clinically normal controls, GII (N = 158), subjects with coronary heart disease without sustained monomorphic ventricular tachycardia (SMVT), and GIII (N = 23), subjects with heart disease and documented SMVT. Conventional variables analyzed from vector magnitude data after averaging to 0.3 μV final noise were obtained by application of each filter to the averaged signal, and evaluated in pairs by numerical comparison and by diagnostic agreement assessment, using conventional and optimized thresholds of normality. Significant differences were found between BI and BD variables in all groups, with diagnostic results showing significant disagreement between both filters [kappa value of 0.61 (P
机译:心室晚期电位是源自受损心肌的低幅度信号,通过ECG滤波和平均在体表上检测到。商业设备中存在的数字过滤器可能会干扰心律失常分层的能力。我们在对时域变量和诊断特性的影响方面将40-Hz BISPEC(BI)和经典的40至250赫兹带传递Butterworth双向(BD)过滤器进行了比较。在横向回顾性的病例调节的病例对照研究中,在信号平均的ECG获取后,将221个具有没有束分支嵌段的窦性节律的受试者,临床正常对照,GII(n = 158),分为三组.GI(n = 40) ,冠心病的受试者没有持续的单声腔室性心动过速(SMVT)和GIII(n = 23),受试者的心脏病和记录的SMVT。通过将每个过滤器应用于平均信号获得从平均到0.3μV最终噪声的矢量幅度数据分析的常规变量通过数值比较和诊断协议评估,通过常规和优化的正常阈值进行评估。在所有组中BI和BD变量之间发现了显着差异,诊断结果显示出在两个过滤器之间的显着分歧[κ值为0.61(p

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号