首页> 外文期刊>BMC Surgery >Design of custom-made navigational template of femoral head and pilot research in total hip resurfacing arthroplasty
【24h】

Design of custom-made navigational template of femoral head and pilot research in total hip resurfacing arthroplasty

机译:总髋关节塑料术中股骨头和试验研究的定制导航模板设计

获取原文
           

摘要

To develop a novel custom-made navigational template for accurate prosthesis implantation in total hip resurfacing arthroplasty (THRA) by computer-aided technology. The template was produced based on data preoperatively acquired from computed tomography (CT) scan. The position of the drill guide was obtained according to the anatomical axis of the femoral neck which was defined by the point of the femoral head center and another point of the femoral neck center. The final direction of the drill guide was confirmed by a valgus angle. The surface of the template was constructed based on the inverse of the femoral neck surface. Then the template was made of acrylate resin by using rapid prototyping (RP) technique. Finally, all the templates were verified in 17 cadavers arranged for THRA and postoperative medical images were employed to evaluate the accuracy and validity of the template. The templates had achieved a high fitting with the femoral neck surface, and there were no guide failures. Postoperative evaluation revealed that the Kirschner-wires pass through the center of the femoral head and femoral neck, presenting a relative expected and acceptable valgus angle to the central axis of the femoral neck. The lateral offset showed the relative valgus angle achieved as expected, the horizontal offset showed that no obvious antero-posterior deviation occured. The comparison between the preoperative Neck-shaft angle (NSA) and the postoperative Stem-shaft angle (SSA) showed there is no significant difference(P??0.05). The novel custom-made navigational template of femoral head can effectively assist surgeons for accurately implanting the femoral head components to the desired position in THRA.
机译:通过计算机辅助技术开发一种新的定制导航模板,用于总髋关节置换术(THRA)的准确假体植入。基于从计算机断层扫描(CT)扫描的数据逐次地产生模板。根据股骨颈部的解剖轴获得钻头的位置,该股骨颈部由股骨头中心的点和股骨颈中心的另一点定义。钻杆的最终方向通过止骨角确认。基于股骨颈表面的逆构成模板的表面。然后通过使用快速原型制作(RP)技术制成模板。最后,所有模板都在安排的17个尸体中验证,采用术后医学图像来评估模板的准确性和有效性。模板已经达到了股骨颈表面的高配件,并且没有导向失败。术后评估显示,Kirschner-丝通过股骨头和股骨颈的中心,向股骨颈的中心轴表示相对预期和可接受的止原角。横向偏移显示,如预期所达到的相对旋流角,水平偏移显示,没有发生明显的缠结后偏差。术前颈轴角度(NSA)与术后茎轴角度(SSA)之间的比较显示没有显着差异(P?>?0.05)。股骨头的新型定制导航模板可以有效地帮助将股骨头部件精确地植入股骨头组件到所需位置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号