The purpose of this study is to compare the predictive accuracy of intraocular lens (IOL) calculations made with partial coherence interferometry (PCI, IOLMaster, version 5) and swept-source optical coherence tomography (SS-OCT, Argos). Axial length (AL), mean keratometry value (K), and anterior chamber depth (ACD) were obtained using PCI and SS-OCT optical biometers. Intraocular lens (IOL) power calculations were made using the Barret-Universal II, Haigis, Hoffer Q, SRK/T, and T2 formulas and compared the predictive accuracy between biometers. In 153 eyes (153 patients), axial length measurements made with PCI (24.65?±?2.35?mm) and SS-OCT (24.62?±?2.29?mm) were significantly different (P?
机译:本研究的目的是比较用部分相干干涉测量(PCI,IOLMASTER,版本5)和扫描源光学相干性断层扫描(SS-OCT,ARGOS)进行的人工晶状体(IOL)计算的预测精度。使用PCI和SS-OCT光学生物夹获得轴向长度(Al),平均腔体值(K)和前腔深度(ACD)。使用麻醉通用II,HAIGIS,Hoffer Q,SRK / T和T2公式进行了人工晶状体(IOL)功率计算,并比较了生物夹之间的预测精度。在153只眼(153名患者)中,用PCI(24.65?±2.35±2.35±2.35±2.35μl)和SS-OCT(24.62?±2.29Ω·毫秒)进行轴向长度测量(P?<0.001)。角膜功率(P?= 0.97)和前腔深度(P?= 0.51)在病灶之间没有显着差异。 PCI或SS-OCT测量的五个IOL功率计算公式之间的平均绝对误差没有显着差异。当Al为24.5-26.0?mm时,源自SS-OCT的平均绝对误差小于所有五个IOL功率计算公式的PCI的平均绝对误差(所有P?<0.05)。总之,PCI和SS-OCT的预测准确性几乎是相同的。然而,在中长的眼睛中,对于IOL计算的SS-OCT的预测精度较高。