A Cautionary Tale: Lack of Consistency in Allele Sizes between Two Laboratories for a Published Multilocus Microsatellite Typing System

机译:告诫故事:两个实验室之间等位基因大小的一致性不足,无法发布已发布的多位点微卫星分型系统

获取原文
       

摘要

For species with low genetic diversity, typing using the differences in PCR fragment length resulting from variations in numbers of short tandem repeats has been shown to provide a high level of discrimination. This technique has been called multilocus microsatellite typing (MLMT) or multiple-locus variable-number tandem repeat analysis, and studies usually employ genetic or sequence analyzers to size PCR fragments to a high degree of precision. We set out to validate one such system that has been developed for Aspergillus fumigatus (H. A. de Valk, J. F. G. M. Meis, I. M. Curfs, K. Muehlethaler, J. W. Mouton, and C. H. W. Klaassen, J. Clin. Microbiol. 43:4112-4120, 2005). The sizes of the alleles were compared both by sequencing and from two genotyping laboratories, where they used capillary electrophoresis (CE) for sizing. Size differences of up to 6 bases were found between the actual sizes reported by sequencing and the sizes reported by CE. In addition, because the two genotyping laboratories used different machines and running conditions, differences of up to 3 bases were identified between them. As the microsatellite markers used differ by repeat units of 3 or 4 bases, it was not possible to assign PCR fragments to the correct alleles without confirming the sizes of a range of alleles by direct sequencing. Lines of best fit were plotted for each CE machine against actual sizes and will therefore enable unsequenced PCR fragments to be assigned to the correct alleles. This study highlights the care required to ensure that an MLMT system undergoes a suitable correction procedure before data can be merged between different laboratories involved in the typing of individual species.
机译:对于遗传多样性较低的物种,使用短串联重复序列数量变化导致的PCR片段长度差异进行分型已显示出很高的区分度。这项技术被称为多基因座微卫星分型(MLMT)或多基因座可变数串联重复分析,并且研究通常使用遗传或序列分析仪对PCR片段进行高精确度测序。我们着手验证一种针对烟曲霉(empergillus fumigatus)开发的系统(HA de Valk,JFGM Meis,IM Curfs,K.Muehlethaler,JW Mouton和CHW Klaassen,J.Clin.Microbiol。 43:4112-4120,2005)。通过测序和两个基因分型实验室对等位基因的大小进行了比较,他们使用毛细管电泳(CE)确定大小。在测序报告的实际大小与CE报告的大小之间发现了最多6个碱基的大小差异。此外,由于两个基因分型实验室使用的机器和运行条件不同,因此在它们之间最多识别出3个碱基的差异。由于使用的微卫星标记的重复单位为3或4个碱基,因此无法通过直接测序来确定等位基因范围的大小而无法将PCR片段分配给正确的等位基因。为每台CE机器绘制了最合适的谱系,并对照实际大小绘制图谱,因此可以将未测序的PCR片段分配给正确的等位基因。这项研究强调了确保MLMT系统经过适当的校正程序之前需要采取的谨慎措施,然后才能在涉及单个物种分类的不同实验室之间合并数据。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号