Evaluation of the MB/BacT system and comparison to the BACTEC 460 system and solid media for isolation of mycobacteria from clinical specimens.

机译:评估MB / BacT系统并与BACTEC 460系统和用于从临床标本中分离分枝杆菌的固体培养基进行比较。

获取原文
获取外文期刊封面目录资料

摘要

The MB/BacT automated system is designed for the isolation of mycobacteria from clinical specimens. It utilizes a colorimetric sensor and reflected light to continuously monitor the CO2 concentration in the culture medium. We compared its performance to that of the BACTEC 12B media for the radiometric BACTEC 460 instrument and that of solid culture media. Respiratory specimens and urine samples were decontaminated with 4% NaOH. The vials of the two instruments were inoculated with 500 microl of sample and two solid egg-based media at 200 microl each. All vials were incubated at 37 degrees C for 6 weeks. A total of 1,078 specimens (633 respiratory specimens, 78 cerebrospinal fluid specimens, 177 other body fluid specimens, 87 urine specimens, and 103 other types of specimens) were cultured in parallel. Mycobacteria could be identified from 73 (6.8%) specimens: 67 M. tuberculosis, 3 M. kansasii, 1 M. xenopi, 1 M. terrae, and 1 mixed M. avium with M. scrofulaceum. Of these, 63 (86.3%) specimens were positive with the MB/BacT system, 67 (91.8%) were positive with the BACTEC 460 instrument, and 58 (79.5%) were positive with the two egg-based media. MB/BacT cultures were positive on average after 17.5 (+/-6.4) days, BACTEC cultures with a growth index of >20 (mean, 200) were positive after 14.3 (+/-8.2) days, and egg-based media were positive after 24.2 (+/-7.5) days. Microorganisms other than mycobacteria contaminated 46 (4.3%) MB/BacT cultures and 31 (2.9%) BACTEC cultures, which had to be discarded. The MB/BacT system is a well-automated system for the detection of M. tuberculosis in clinical specimens without using radioactive reagents. Further trials are required to determine whether it is suitable for the culture of nontuberculous mycobacteria.
机译:MB / BacT自动化系统设计用于从临床标本中分离分枝杆菌。它利用比色传感器和反射光来连续监测培养基中的CO2浓度。我们将其性能与辐射型BACTEC 460仪器的BACTEC 12B培养基的性能以及固体培养基的性能进行了比较。呼吸道标本和尿液标本用4%NaOH消毒。用500微升样品和两种固体蛋基培养基分别以200微升接种两种仪器的小瓶。将所有小瓶在37℃下孵育6周。平行培养总共1078个标本(633个呼吸道标本,78个脑脊液标本,177个其他体液标本,87个尿液标本和103个其他类型的标本)。可以从73个(6.8%)标本中鉴定出分枝杆菌:67 M.结核病,3 M. kansasii,1 M. xenopi,1 M. terrae和1个鸟笼分枝杆菌与s分支杆菌。其中,使用MB / BacT系统的63个样本(86.3%)为阳性,使用BACTEC 460仪器的67个样本(91.8%)为阳性,使用两种基于蛋的培养基中的58个样本(79.5%)为阳性。 MB / BacT培养物在17.5(+/- 6.4)天后平均呈阳性,生长指数> 20(平均200)的BACTEC培养物在14.3(+/- 8.2)天后呈阳性,而卵基培养基为24.2(+/- 7.5)天后为阳性。除分枝杆菌外的微生物污染了46(4.3%)MB / BacT培养物和31(2.9%)BACTEC培养物,必须将其丢弃。 MB / BacT系统是一种高度自动化的系统,无需使用放射性试剂即可检测临床标本中的结核分枝杆菌。需要进一步的试验来确定它是否适合于非结核分枝杆菌的培养。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号