The present study investigated the feasibility of automating the specimen-pipetting component of sample preparation in the LCx Chlamydia assay (LCx-CT assay; Abbott Laboratories, Chicago, Ill.) by using a commercially available liquid-handling system (Tecan Genesis RSP100; Tecan Inc., Research Triangle Park, N.C.). The Tecan instrument proved to be comparable in both precision and accuracy to a manual multipipettor (Eppendorf model 4850; Eppendorf Scientific, Westbury, N.Y.). The Tecan instrument was extensively checked for evidence of specimen-to-specimen transfer, and no level of contamination sufficient to generate a signal above the background in the LCx-CT assay was detected. Finally, pipetting speed was significantly improved by using the Tecan instrument. A mean time of 2.5 min was required to pipette a complete LCx-CT assay carousel (20 samples and 4 controls) with the Tecan instrument, whereas 8.4 min was required to pipette a comparable number of samples manually (P < 0.001).
机译:本研究调查了使用市售液体处理系统(Tecan Genesis RSP100; Tecan)在LCx衣原体测定(LCx-CT测定;雅培,伊利诺伊州芝加哥)中自动进行样品制备的样品移液组件的可行性。公司,北卡罗莱纳州研究三角公园)。事实证明,Tecan仪器在精度和准确性上都可与手动多管移液器(Eppendorf 4850型; Eppendorf Scientific,纽约州韦斯特伯里)相媲美。对Tecan仪器进行了广泛的检查,以检查是否存在样品到样品的转移,并且在LCx-CT分析中未检测到足以产生高于本底信号的污染水平。最后,使用Tecan仪器可以显着提高移液速度。用Tecan仪器吸取完整的LCx-CT分析转盘(20个样品和4个对照)需要平均2.5分钟,而手动吸取相当数量的样品( P <0.001)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号