首页> 外文期刊>Journal of Clinical Microbiology >Comparison of the Binax Legionella Urinary Antigen Enzyme Immunoassay (EIA) with the Biotest Legionella Urin Antigen EIA for Detection of Legionella Antigen in both Concentrated and Nonconcentrated Urine Samples
【24h】

Comparison of the Binax Legionella Urinary Antigen Enzyme Immunoassay (EIA) with the Biotest Legionella Urin Antigen EIA for Detection of Legionella Antigen in both Concentrated and Nonconcentrated Urine Samples

机译:Binax军团菌尿液抗原酶联免疫分析(EIA)与Biotest军团菌尿液抗原EIA的比较,可用于检测浓缩和非浓缩尿液样品中的军团菌抗原

获取原文
           

摘要

We evaluated a newly commercial enzyme immunoassay (EIA) (BiotestLegionella Urin Antigen EIA; Biotest AG, Dreieich, Germany) for detection of antigens of all Legionella pneumophilaserogroups with a relatively wide spectrum of cross-reactivity as well as antigens of other Legionella spp. by comparing its sensitivity and specificity with those of an EIA for detection ofL. pneumophila serogroup 1 antigen (Legionella urinary antigen EIA; Binax, Portland, Maine). Both tests were performed with both concentrated and nonconcentrated urine samples. We also evaluated the capabilities of both EIAs to detect extracted soluble antigens of American Type Culture Collection (ATCC) Legionella strains (L. pneumophila serogroups 1 to 14, L. bozemanii, and L. longbeachae). The sensitivity of the Biotest EIA was 66.66% in nonconcentrated urine and 86.66% in concentrated urine. The sensitivity of the Binax EIA was 63.76% and 88.88% in nonconcentrated and concentrated urine, respectively. The specificity was 100% in nonconcentrated and concentrated urine for both assays. The Binax EIA and Biotest EIA detected extracted soluble antigens of L. pneumophila serogroups 1 to 14 and L. bozemanii ATCC strains. The cross-reactions observed with the Binax EIA were probably due to common epitopes directly related to lipopolysaccharide. Further studies are required to determine the usefulness of the Binax EIA for detection of urinary antigens fromLegionella species and serogroups other than L. pneumophila serogroup 1. The Biotest EIA proved to be as rapid, sensitive, and specific as the Binax EIA for the diagnosis of legionellosis. Concentration of antigen present in urine increased the sensitivities of both techniques with no reduction in specificity.
机译:我们评估了一种新的商业酶免疫测定(EIA)(Biotest Legionella 尿素抗原EIA; Biotest AG,Dreieich,德国),用于检测所有嗜肺军团杆菌(Legionella pneumophila)血清群的抗原广泛的交叉反应性以及其他 Legionella spp的抗原。通过将其灵敏度和特异性与EIA检测Eem的灵敏度和特异性进行比较。肺炎血清群1抗原(尿抗原EIA; Binax,缅因州波特兰)。两种测试均针对浓缩尿液样本和非浓缩尿液样本进行。我们还评估了两种EIA检测美国典型培养物保藏中心(ATCC)军团菌(em> Legionella 菌株( L。pneumophila 血清群1至14, L)的提取可溶性抗原的能力。 。bozemanii longbeachae )。 Biotest EIA的灵敏度在非浓缩尿液中为66.66%,在浓缩尿液中为86.66%。 Binax EIA在非浓缩尿液和浓缩尿液中的敏感性分别为63.76%和88.88%。两种检测在非浓缩尿液和浓缩尿液中的特异性均为100%。 Binax EIA和Biotest EIA检测到提取的 L可溶性抗原。肺炎血清群1至14和 L。 bozemanii ATCC菌株。 Binax EIA观察到的交叉反应可能是由于与脂多糖直接相关的常见表位所致。需要进一步的研究来确定Binax EIA对检测来自 Legionella 物种和除 L之外的血清群的尿抗原的有用性。肺炎血清群1. Biotest EIA与Binax EIA一样快速,灵敏且特异,可用于军团病的诊断。尿液中存在的抗原浓度增加了这两种技术的敏感性,而特异性没有降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号