The Impact of Lens Opacity on SD-OCT Retinal Nerve Fiber Layer and Bruch's Membrane Opening Measurements Using the Anatomical Positioning System (APS)

机译:晶状体不透明度对SD-OCT视网膜神经纤维层的影响以及使用解剖定位系统(APS)进行的Bruch膜开口测量

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Purpose: To evaluate the impact of lens opacity on retinal nerve fiber layer thickness (RNFLT) and Bruch's membrane opening (BMO) measurements. Methods: Fifty-nine randomly selected patients without any other relevant ocular pathology undergoing elective routine cataract surgery in two specialized eye clinics were enrolled. RNFLT, BMO area, and BMO minimum rim width (BMO-MRW) were assessed with the Heidelberg Engineering Spectralis OCT using the anatomical positioning system (APS) prior to and 1 day after cataract surgery using a ring scan at different eccentricities of the disc (3.5, 4.1 and 4.7 mm). Lens opacity was quantified using densitometry based on Scheimpflug images (Oculus Pentacam AXL). Results: RNFLT, BMO area, and BMO-MRW were virtually identical before and following removal of the cataractous lens. This held when assessed overall, within the six sectors for the 3.5-mm scan, or at any other eccentricity. Baseline RNFLT was not associated with lens opacity. Conclusions: Using the APS, RNFLT remained unchanged following cataract surgery, contrary to results reported by previous studies. Our results imply that the APS may have contributed to more precise spectral-domain optical coherence measurements, minimizing the influence of cataract on RNFLT and BMO assessments in our cohort.
机译:目的:评估晶状体混浊度对视网膜神经纤维层厚度(RNFLT)和Bruch膜开口(BMO)测量的影响。方法:招募了在两家专门的眼科诊所接受选择性择期白内障手术的59例随机选择的无其他相关眼部病理的患者。在白内障手术前和手术后1天,使用环形定位系统在不同的偏心距盘上进行环形扫描,使用Heidelberg Engineering Spectralis OCT使用解剖定位系统(APS)对RNFLT,BMO面积和BMO最小边缘宽度(BMO-MRW)进行了评估( 3.5、4.1和4.7毫米)。使用基于Scheimpflug图像(Oculus Pentacam AXL)的光密度法对晶状体的不透明度进行定量。结果:在去除白内障晶状体之前和之后,RNFLT,BMO面积和BMO-MRW几乎相同。总体评估时,在3.5毫米扫描的六个扇区内,或在任何其他偏心率下,此值均保持不变。基线RNFLT与晶状体混浊无关。结论:使用APS,白内障手术后RNFLT保持不变,这与先前研究报告的结果相反。我们的结果表明,APS可能有助于更精确的光谱域光学相干测量,从而将白内障对我们队列中RNFLT和BMO评估的影响降至最低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号