Morphology and Repeatability of Automated Perimetry using Stimulus Size VI: A Comparison with Sizes III and V

机译:使用刺激尺寸VI进行自动视野测量的形态和可重复性:与尺寸III和V的比较

获取原文
获取外文期刊封面目录资料

摘要

? Purpose: ?To describe the visual field of normal subjects using Humphrey perimetry with size VI stimuli and to compare size VI thresholds and retest variability with perimetry using sizes III and V stimuli. ? Methods: ?We tested one eye of 60 normal subjects with the Humphrey full threshold algorithm using sizes V (1.72°) and VI (3.44°) stimuli and with SITA-Standard? using size III stimuli (0.43°). The patients were retested 1-4 weeks later. We compared the mean scores, concentric zones, and point-wise sensitivities among the sizes and their retest variability. Repeated measures ANOVA on Ranks was performed with the dependent variable as sensitivity (dB) of average sensitivity of each eccentric zone. ? Results: ?The mean sensitivities were size III: 30.2 ± 1.1; V: 34.4 ± 1.0 and VI 36.0 ± 1.0 (p 0.001 with all Tukey post hoc paired comparisons significant). Significant differences between the groups were also present for each eccentric zone except 0° for size V vs. size VI. The mean difference on retest across test locations was 1.55 dB ± 1.37 for size III, 1.36 dB ± 1.23 for size V, and 1.27 dB ± 1.15 for size VI. The difference in variability between size III, V and size VI increased with eccentricity, with size III increasing more than the larger stimulus sizes but statistical significance for this difference was not reached. (Figure). ? Conclusions: ?The retest variability in normal subjects is slightly less for sizes V and VI full threshold testing compared with size III SITA-Standard? results and for larger sizes there is a minimal increase with eccentricity. ?View OriginalDownload SlideView OriginalDownload Slide ? Keywords: perimetry ? visual fields?.
机译:?目的:To描述使用汉弗莱视野测量法和VI刺激物对正常受试者的视野,并比较视野大小VI阈值和使用视野测量法用III和V刺激物重新测试变异性。 ?方法:我们使用Humphrey全阈值算法使用大小为V(1.72°)和VI(3.44°)的刺激以及SITA-Standard测试了60名正常受试者的一只眼睛。使用III号刺激(0.43°)。 1-4周后重新检查患者。我们比较了平均得分,同心区域和大小之间的逐点敏感性及其重测变异性。使用因变量作为每个偏心区平均灵敏度的灵敏度(dB)进行了Rank上的重复测量方差分析。 ?结果:平均敏感度为III级:30.2±1.1; V:34.4±1.0和VI 36.0±1.0(p <0.001,所有Tukey事后配对比较显着)。对于每个偏心区域,组之间也存在显着差异,除了尺寸V与尺寸VI的0°不同。跨测试位置的重新测试的平均差异对于尺寸III为1.55 dB±1.37,对于尺寸V为1.36 dB±1.23,对于尺寸VI为1.27 dB±1.15。大小III,V和大小VI之间的变异性差异随偏心率而增加,其中大小III的增加幅度大于较大的刺激大小,但未达到该差异的统计意义。 (数字)。 ?结论:?与III型SITA标准相比,V型和VI型全阈值测试的正常受试者的重测变异性略低?结果,对于较大的尺寸,偏心率的增加很小。查看原始下载幻灯片查看原始下载幻灯片关键词:视野检查视野?

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号