Discordance in the minimal inhibitory concentrations of ertapenem for Enterobacter cloacae: Vitek 2 system versus Etest and agar dilution methods

机译:阴沟肠杆菌厄他培南的最低抑菌浓度不一致:Vitek 2系统与Etest和琼脂稀释方法

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Our objective was to compare the ertapenem minimal inhibitory concentrations (MICs) for Enterobacter cloacae isolates categorized intermediate or resistant to ertapenem when measured with the Vitek 2 system, with the MICs for these isolates when measured by two methods performed in agar medium: the Etest and agar plate dilution method (APDM). Overall, 50 E. cloacae isolates were included in the study. The mean MIC of ertapenem was 2.92+/-1.77@mg/ml according to the Vitek 2 system, 0.94+/-0.84@mg/ml according to the Etest strips, and 0.93+/-0.62@mg/ml according to the APDM. Furthermore, the MICs determined by the Vitek 2 system were higher than the MICs determined by the two other methods for 96% of strains. Lastly, according to the Etest strips and APDM, 42% of E. cloacae were susceptible to ertapenem. No carbapenemase was identified by the screening method used. Using the Vitek 2 system to determine ertapenem MICs for E. cloacae can have potential consequences in terms of additional carbapenemase-detecting tests and antimicrobial therapy. It would be interesting to determine if the Vitek 2 system is more effective for the detection of carbapenemase producers with low-level carbapenem resistance than the two methods performed in agar medium.
机译:我们的目的是比较使用Vitek 2系统进行测量时,阴沟肠杆菌分离株的中等抑或ertapenem耐药菌的最小抑菌浓度(MICs)和通过两种方法在琼脂培养基中进行测定时对这些分离株的MICs的MICs进行比较:琼脂平板稀释法(APDM)。总体而言,研究中包括了50种阴沟肠杆菌。根据Vitek 2系统,ertapenem的平均MIC为2.92+/-1.77@mg/ml,根据Etest试纸的平均MIC为0.94+/-0.84@mg/ml,根据Etest试纸的平均MIC为0.93+/-0.62@mg/ml APDM。此外,对于96%的菌株,通过Vitek 2系统测定的MIC高于通过其他两种方法测定的MIC。最后,根据Etest试纸和APDM,阴沟肠杆菌中有42%对厄他培南敏感。通过所用的筛选方法未鉴定出碳青霉烯酶。使用Vitek 2系统确定阴沟肠杆菌的ertapenem MIC可能会对其他碳青霉烯酶检测和抗菌治疗产生潜在的影响。确定Vitek 2系统是否比在琼脂培养基中进行的两种方法更有效地检测具有低水平碳青霉烯耐药性的碳青霉烯酶生产者。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号