首页> 外文期刊>Indian Journal of Science and Technology >Comparative Analysis and Accuracy of a Devised Automated Non Invasive Blood Pressure Monitor Based on Oscillometric Method
【24h】

Comparative Analysis and Accuracy of a Devised Automated Non Invasive Blood Pressure Monitor Based on Oscillometric Method

机译:一种基于示波法的自动无创血压监测仪的比较分析与精度

获取原文
获取外文期刊封面目录资料

摘要

Objectives: This paper focuses on the evaluation of accuracy and reliability of previously designed oscillometric based Non Invasive Blood Pressure (NIBP) instrument by means of a comparative analysis with the standard cuff mercury sphygmomanometer based on auscultation. Methods/Statistical Analysis: Thirty healthy subjects between the ages of 18-23 were examined by successive application of both devices having identical cuffs and site of measurement. The analysis of means and bias values demonstrated that our designed oscillometric NIBP monitor under-read all blood pressure values compared to mercury manometer. Findings: The mean oscillometric NIBP readings (Systolic Blood Pressure (SBP) 119.65 ± 8.25, Diastolic Blood Pressure (DBP) 76.82 ± 6.12 and Mean Arterial Pressure (MAP) 91.1 ± 4.76) were significantly lower than its auscultatory counterpart (SBP 127.03 ± 8.7, DBP 81.4 ± 6.2 and MAP 96.6 ± 4.95). F-test showed that the devised prototype is precise and comparable to a reference mercury manometer. A p-value of <0.05 was recognized to be statistically considerable. Conclusively, the designed NIBP monitoring prototype exhibited a tendency to measure lower blood pressure values than the auscultatory sphygmomanometer; but both devices were well comparable. The difference found, was statistically significant but below the level of physiological significance. Application/Improvements: This oscillometric NIBP prototype may be used for training as well as learning objectives. Furthermore, this prototype can be improved by Arduino based microcontroller and kit in the future studies.
机译:目的:本文重点通过基于听诊的标准袖带水银血压计的对比分析,评估先前设计的基于示波器的无创血压(NIBP)仪器的准确性和可靠性。方法/统计分析:通过连续使用两个具有相同袖口和测量部位的装置,对30名18-23岁的健康受试者进行了检查。对均值和偏差值的分析表明,与汞压力计相比,我们设计的示波NIBP监测器无法读取所有血压值。结果:示波平均NIBP读数(收缩压(SBP)119.65±8.25,舒张压(DBP)76.82±6.12和平均动脉压(MAP)91.1±4.76)显着低于听诊时的相对血压(SBP 127.03±8.7) ,DBP 81.4±6.2和MAP 96.6±4.95)。 F检验表明,设计的原型精确且可与参考汞压力计媲美。 P值<0.05被认为具有统计学意义。结论是,设计的NIBP监测原型具有比听诊血压计更低的血压测量趋势。但两种设备的可比性都很好。发现的差异具有统计学显着性,但低于生理学显着性水平。应用/改进:该示波NIBP原型可用于培训和学习目标。此外,在将来的研究中,可以通过基于Arduino的微控制器和套件来改进此原型。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号