This paper presents the recycling possibility of irradiated alanine dosimeter via annealing methods. The alanine samples were irradiated under 10 MeV electron beam with the doses of 10, 30, and 60 kGy. The samples were stored in an oven with the different temperature and time. The fading effects of the EPR signal of the alanine samples were investigated, and the optimum time and temperature were found. Furthermore, the calibration curves have been drawn for the annealed samples. The stability of annealed alanine has also been examined. The results show that the optimum annealing time and temperature are 30 min and 200 °C, respectively. On the other hand, it was obvious that the obtained results were independent of the radiation dose.
机译:本文介绍了通过退火方法辐照的丙氨酸剂量计的再循环可能性。在10MeV电子束下以10、30和60kGy的剂量照射丙氨酸样品。将样品存储在具有不同温度和时间的烤箱中。研究了丙氨酸样品的EPR信号的衰落效应,并找到了最佳时间和温度。此外,已绘制了退火样品的校准曲线。还研究了退火的丙氨酸的稳定性。结果表明,最佳退火时间和温度分别为30分钟和200°C。另一方面,很明显,所获得的结果与辐射剂量无关。