首页> 外文期刊>Applied Microbiology >Procedure for the Evaluation of the Virucidal Effectiveness of an Ethylene Oxide Gas Sterilizer
【24h】

Procedure for the Evaluation of the Virucidal Effectiveness of an Ethylene Oxide Gas Sterilizer

机译:环氧乙烷气体灭菌器杀毒效果评估程序

获取原文
           

摘要

A quantitative, reproducible method was developed for the evaluation of the virucidal activity of test gases. Using this method, we determined the virucidal effectiveness of a Steri-Vac ethylene oxide gas sterilizer. Wool gabardine material was exposed to high concentrations of herpes simplex, vaccinia, parainfluenza, or polio viruses and was processed through the sterilizer. Two time-temperature cycles of the machine, 29 C for 180 min and 60 C for 48 min, were used in separate experiments. The viruses were exposed to the gas when freshly pipetted onto the fabric or when pipetted on the material and allowed to dry 16 to 24 hr. In two experiments carried out under each condition, the virus titers were reduced by the sterilization process to less than detectable limits. These titer reductions were for the herpes virus ≥ 2.7 to 5.0 log, for vaccinia virus ≥ 4.0 to 6.1 log, for parainfluenza virus ≥ 1.8 to 4.9 log, and for poliovirus ≥ 4.9 to 7.7 log. The observed reductions in virus titers were the same whether the virus-contaminated fabrics were sealed in polyethylene packages or held in open petri dishes during exposure to ethylene oxide.
机译:开发了一种定量,可重现的方法来评估测试气体的杀病毒活性。使用这种方法,我们确定了Steri-Vac环氧乙烷气体灭菌器的杀菌效力。将羊毛华达呢丁材料暴露于高浓度的单纯疱疹,牛痘,副流感或小儿麻痹症病毒中,并通过灭菌器进行处理。在单独的实验中使用了机器的两个时间-温度循环,即29°C持续180分钟和60°C持续48分钟。新鲜吸取到织物上或吸取到材料上时,病毒会暴露在气体中,并干燥16至24小时。在每种条件下进行的两次实验中,通过灭菌过程将病毒滴度降低到可检测的极限以下。滴度下降的幅度是:疱疹病毒≥2.7至5.0 log,牛痘病毒≥4.0至6.1 log,副流感病毒≥1.8至4.9 log,脊髓灰质炎病毒≥4.9至7.7 log。无论在暴露于环氧乙烷的过程中,被病毒污染的织物是密封在聚乙烯包装中还是放在开放的培养皿中,病毒滴度的降低都是相同的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号