Analysis of the Physicochemical State of C-Reactive Protein in Different Preparations Including 2 Certified Reference Materials

机译:包括2种经过认证的参考物质在内的不同制剂中C反应蛋白的理化状态分析

获取原文
           

摘要

BACKGROUND: Standardization of clinical measurements of C-reactive protein (CRP) is based on the availability of reference materials. The serum protein reference material ERM-DA470 was used as a calibrant for various commercially available immunoassays but has now been exhausted. The recently released ERM-DA470k/IFCC was intended to fully replace ERM-DA470. However, the new material was not suited for the certification of CRP because of a bias introduced by the lyophilization process that caused loss of about 20% of CRP measurable by routine immunoassays, compared with the nonlyophilized material that was stored in a liquid frozen state.METHODS: We investigated the physicochemical state of CRP in a set of 4 lyophilized and 2 nonlyophilized serum-based CRP-containing materials by semi-native gel electrophoresis, Western blotting, and gel filtration.RESULTS: We detected a monomeric form of CRP (mCRP) in lyophilized materials at a concentration significantly higher than seen in the materials not subjected to lyophilization. Different reconstitution protocols led to variations of the monomeric CRP fraction found in reconstituted, previously lyophilized material.CONCLUSIONS: Most of the 20% loss in measured CRP after lyophilization of the material can be accounted for by the dissociation of natively pentameric CRP into subunits. The observed dissociation results from lyophilization and subsequent reconstitution of the material at insufficient concentration levels of calcium ions. In the presence of various protein forms, differences in antibody specificity and reactivity between immunoassays and alterations of stoichiometry of antigen–antibody interactions can contribute to the divergence of the measured values.
机译:背景:C反应蛋白(CRP)的临床测量标准化基于参考材料的可用性。血清蛋白参考物质ERM-DA470被用作各种商业化免疫测定的校准物,但现在已经用尽。最近发布的ERM-DA470k / IFCC旨在完全替代ERM-DA470。但是,这种新材料不适合CRP认证,因为冻干过程引入了偏差,与以液态冷冻状态存储的非冻干材料相比,常规免疫测定可测量到约20%的CRP损失。方法:我们通过半天然凝胶电泳,蛋白质印迹和凝胶过滤研究了4种冻干和2种非冻干血清基含CRP材料中CRP的理化状态。结果:我们检测到CRP(mCRP)的单体形式)在冻干材料中的浓度显着高于未进行冻干的材料中的浓度。不同的重组方案导致重组的,先前冻干的材料中发现的单体CRP分数发生变化。观察到的离解是由于冻干和随后在钙离子浓度不足的情况下材料的重建所致。在存在各种蛋白质形式的情况下,免疫测定之间抗体特异性和反应性的差异以及抗原-抗体相互作用的化学计量的改变可能会导致测量值的差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号