首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >A case of sulfhemoglobinemia and emergency measurement of sulfhemoglobin with an OSM3 CO-oximeter
【24h】

A case of sulfhemoglobinemia and emergency measurement of sulfhemoglobin with an OSM3 CO-oximeter

机译:一例血红蛋白血红蛋白血症和使用OSM3 CO血氧仪对血红蛋白血红蛋白进行紧急检测

获取原文
       

摘要

We describe a case of sulfhemoglobinemia associated with toxic paint ingestion. Blood gases, oxygen content, and fractional hemoglobin derivatives were assayed with Radiometer 520 and OSM3 instruments. Although the CO-oximeters indicated the presence of sulfhemoglobin (SulfHb), the results were not quantitative. An OSM3 service software program was activated to obtain the actual concentrations of the hemoglobin fractions. Subsequently, we evaluated the performance of the OSM3 service program for the analysis of SulfHb by performing precision studies and comparing OSM3 results with those of an AVL 912 CO-oximeter. Retrospectively, we determined that the patient’s specimens contained 6% SulfHb. There was an obvious deviation between standard OSM3 oxyhemoglobin fraction measurements and those obtained by using its service program—the effect of a high SulfHb content.
机译:我们描述了与有毒涂料摄入有关的血红蛋白血红蛋白病例。用Radiometer 520和OSM3仪器测定血气,氧气含量和部分血红蛋白衍生物。尽管CO血氧饱和度计表明存在硫血红蛋白(SulfHb),但结果不是定量的。激活OSM3服务软件程序以获得血红蛋白组分的实际浓度。随后,我们通过进行精密度研究并将OSM3结果与AVL 912 CO血氧仪的结果进行了比较,评估了OSM3服务程序用于分析SulfHb的性能。回顾性地,我们确定该患者的标本中含有6%的SulfHb。标准OSM3氧合血红蛋白分数测量值与使用其服务程序获得的分数之间存在明显的偏差-高SulfHb含量的影响。