Comparative evaluation of three assay systems for automated determination of hemoglobin A1c

机译:三种测定系统自动测定血红蛋白A1c的比较评估

获取原文
       

摘要

We evaluated three newly introduced systems for automated determinations of hemoglobin (Hb) A1c, which allow the processing of large amounts of samples in a routine clinical laboratory. We compared these methods—the Variant HPLC, the Hi-Auto A1c analyzer system, and the Roche immunoassay—with the Diamat HPLC system. All showed good precision and good concordance with the Diamat HPLC. The reference range for Hb A1c has to be determined by the laboratory for each assay system. Interference study showed no statistically significant influence of anemia, polycythemia, rheumatoid factor, or chronic hemodialysis, although individual Hb A1c values can be influenced by polycythemia (when measured with the Hi-Auto A1c analyzer) and by chronic hemodialysis (when measured with the Variant HPLC). HPLC was not suitable for measuring Hb A1c in the examined cases of hemoglobin variants; assaying fructosamine seems to be better for monitoring these patients.
机译:我们评估了三个新引入的系统,用于自动测定血红蛋白(Hb)A1c,从而可以在常规临床实验室中处理大量样品。我们将这些方法(变体HPLC,Hi-Auto A1c分析仪系统和罗氏免疫测定法)与Diamat HPLC系统进行了比较。与Diamat HPLC均显示出良好的精密度和一致性。 Hb A1c的参考范围必须由实验室针对每个分析系统确定。干扰研究显示,尽管个别Hb A1c值可能受红细胞增多症(使用Hi-Auto A1c分析仪测量)和慢性血液透析(当使用Variant进行测量时)的影响,但贫血,红细胞增多症,类风湿因子或慢性血液透析的统计学影响均无统计学意义。 HPLC)。在检查的血红蛋白变异病例中,HPLC不适合测量Hb A1c。测定果糖胺似乎更好地监测了这些患者。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号