Frequency of Extreme Differences and Clinical Performance of Glucose Concentration Measurements Judged from 21 000 Duplicate Measurements

机译:从21000次重复测量中判断出的极端差异的频率和血糖浓度测量的临床表现

获取原文
   

获取外文期刊封面封底 >>

       

摘要

To the Editor:Duplicate measurements of patient samples are not common in routine clinical chemistry. The performance of measurements is continuously monitored by the internal quality control (IQC),1 which is not designed to find occasional dropouts. Therefore, the frequency of such errors is not usually known. In the present study, we attempted to estimate the error frequency at commonly suggested imprecision targets by evaluating the difference between 2 adjacent repeated measurements. The frequency is expected to be very low, and therefore the study of this problem required a large number of samples.We measured glucose concentration over a 6-month period in samples obtained from an unselected population of hospitalized patients and outpatients. We used 3 Dimension Vista 1500 analyzers connected to StreamLAB (Siemens Healthcare Diagnostics), in accordance with the manufacturer's instructions. To include measured values in our study, we required that glucose assay performance comply with the RiLiBAEK (Richtlinien der Bundes?rztekammer) IQC limits (1) over the measuring interval of 1 mg/dL (0.06 mmol/L) to 500 mg/dL (27.75 mmol/L).We measured 21 653 samples twice, with each sample randomly allocated to one of the instruments by the laboratory automation system. The repeat measurement, which …
机译:致编辑:在常规临床化学中,重复测量患者样品并不常见。内部质量控制(IQC)[1]持续监控测量的性能,该质量控制并非旨在发现偶尔的脱落。因此,这种错误的发生频率通常是未知的。在本研究中,我们尝试通过评估两个相邻重复测量之间的差异来估计通常建议的不精确目标的错误频率。预计频率会非常低,因此研究此问题需要大量样本。我们测量了六个月期间从未经选择的住院患者和门诊病人中获得的样本中的葡萄糖浓度。我们按照制造商的说明使用了连接到StreamLAB(Siemens Healthcare Diagnostics)的3维Vista 1500分析仪。为了将测量值包括在我们的研究中,我们要求葡萄糖测定性能在1 mg / dL(0.06 mmol / L)至500 mg / dL的测量间隔内符合RiLiBAEK(Richtlinien der Bundes?rztekammer)IQC限值(1)。 (27.75 mmol / L)。我们对21 653个样品进行了两次测量,每个样品均由实验室自动化系统随机分配给其中一种仪器。重复测量,…

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号