Development and analytical performance of a functional assay for fibrinogen on the Du Pont aca analyzer.

机译:在杜邦aca分析仪上进行纤维蛋白原功能测定的开发和分析性能。

获取原文
           

摘要

A method for determination of functional fibrinogen has been developed for the Du Pont aca discrete clinical analyzer. This fully automated assay is based on the rate of fibrin turbidity formation when thrombin is added to the test sample. The rate of this reaction is enhanced by added calcium chloride and dextran. The assay range is from 0.50 to 8.00 g/L. Within-run reproducibility studies at the medical decision level of 1.0 g/L showed a CV of 2.5%. There was no interference by heparin up to 10 USP units per milliliter of plasma.
机译:已经为Du Pont aca离散临床分析仪开发了一种测定功能性纤维蛋白原的方法。这种全自动测定是基于将凝血酶添加到测试样品中时血纤蛋白浊度的形成速率。加入氯化钙和葡聚糖可提高反应速度。测定范围是0.50至8.00g / L。在医学决策水平为1.0 g / L的运行内可重复性研究中,CV为2.5%。每毫升血浆中高达10 USP单位的肝素没有干扰。