摘要

We compared a modified enzymic continuous-flow method for serum triglycerides, in which a microbial lipase is used without protease, to an extraction-fluorometric method. Results (for total--blank) correlated well with those by the comparison procedure: enzymic = 1.002 fluorometric--14 mg/liter; r = 0.9968. Serum blanks were measured for each sample in both methods, and the concentration of free glycerol measured by the enzymic method was shown to be as much as 15 to 71% of the total triglyceride result in 7% of the samples analyzed. The dual-channel continuous-flow system provides automatic free-glycerol blank subtraction for the enzymic method.
机译:我们将血清甘油三酸酯的改进的酶促连续流方法与提取荧光法进行了比较,在该方法中,使用了不添加蛋白酶的微生物脂肪酶。结果(总空白)与比较程序的相关性很好:酶学= 1.002荧光--14 mg / L; r = 0.9968。在两种方法中都对每种样品进行了血清空白测定,在7%的分析样品中,通过酶法测定的游离甘油浓度显示为甘油三酯总量的15%至71%。双通道连续流系统为酶法提供了自动的游离甘油空白扣除。