Ion-selective membrane electrode for rapid automated determinations of total carbon dioxide.

机译:离子选择膜电极可快速自动测定总二氧化碳。

获取原文
获取外文期刊封面目录资料

摘要

Previously described ion-selective membrane electrodes useful for determination of total carbon dioxide in biological fluids in automated analyzers show interference from several sources--e.g., fatty acids, keto acids, salicylate, heparin. Judicious selection of membrane components has produced a membrane with superior performance characteristics: short conditioning time, long lifetime in storage, rapid and stable response, low drift, and significantly less susceptibility to interference than other electrodes thus far reported. Samples can be analyzed at 240 samples or more per hour and results correlate well with those by, e.g., the Technicon SMAC II method for total carbon dioxide. Mounted on the Technicon RA-1000, the carbon dioxide sensor is arranged in tandem with sodium and potassium ion-selective electrodes. When all three species are measured in the same buffered sample stream, the sampling rate thus becomes 720 tests per hour, done on 25-microL samples.
机译:先前描述的可用于测定自动分析仪中生物流体中总二氧化碳的离子选择膜电极表现出来自多种来源的干扰-例如脂肪酸,酮酸,水杨酸盐,肝素。与迄今报道的其他电极相比,明智地选择膜组件可以生产出一种性能优异的膜:调节时间短,存储寿命长,响应快速稳定,漂移小,对干扰的敏感性明显降低。每小时可以分析240个或更多样本,并且其结果与通过Technicon SMAC II方法测得的二氧化碳总量具有很好的相关性。二氧化碳传感器安装在Technicon RA-1000上,与钠和钾离子选择电极串联安装。当在相同的缓冲样品流中测量所有三种物质时,采样率将变为每小时720次测试,对25微升样品进行。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号