首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Immunoturbidimetric assay for alpha 1-acid glycoprotein in serum with a centrifugal analyzer.
【24h】

Immunoturbidimetric assay for alpha 1-acid glycoprotein in serum with a centrifugal analyzer.

机译:用离心分析仪进行免疫比浊法测定血清中的α1-酸糖蛋白。

获取原文
           

摘要

This is a simple immunoturbidimetric method for the rapid assessment of alpha 1-acid glycoprotein in serum with the "Flexigem" microcentrifugal analyzer. Results obtained for a pediatric population (n = 95) agree with those by radial immunodiffusion (x) and yield the regression equation: y = 1.4x + 0.017 (r = 0.98). Results are printed automatically after a procedure that takes only about 15 min. Within-run precision (CV) ranged from 2.2 to 2.5% for concentrations of 0.35 to 2.4 g/L. Day-to-day CV was 4.25% for an average concentration of 0.9 g/L.
机译:这是一种使用“ Flexigem”微量离心分析仪快速评估血清中α1-酸糖蛋白的简单免疫比浊法。儿科人群(n = 95)获得的结果与放射免疫扩散(x)结果相符,得出回归方程:y = 1.4x + 0.017(r = 0.98)。只需大约15分钟的操作,结果就会自动打印出来。对于0.35至2.4 g / L的浓度,批内精密度(CV)为2.2%至2.5%。每日CV为4.25%,平均浓度为0.9 g / L。