首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Evaluation for Four Reagent Kits and Two Flame Photometers Used to Determine Sodium and Potassium in Serum
【24h】

Evaluation for Four Reagent Kits and Two Flame Photometers Used to Determine Sodium and Potassium in Serum

机译:评估用于测定血清中钠和钾的四个试剂盒和两个火焰光度计

获取原文
           

摘要

The Stanbio and Medi-Chem sodium and potassium kits were evaluated and the results compared with those obtained with the Coleman Jr. Model 21, NIL 4-7000, and Technicon flame photometers. Day-to-day reproducibility studies and data based on comparative studies with sera from patients revealed that the Stanbio and Medi-Chem sodium kits and the Medi-Chem potassium kit give diagnostically unsatisfactory results. Results with the Stanbio potassium kit were reproducible, but further improvement is required, as indicated by nonspecificity in the analyses of patient sera. Instability of the Coleman Jr. flame photometer made it difficult to obtain acceptable data in these tests, particularly for sodium, while the NIL flame photometer had good precision and accuracy.
机译:对Stanbio和Medi-Chem钠和钾试剂盒进行了评估,并将结果与​​使用Coleman Jr. Model 21,NIL 4-7000和Technicon火焰光度计获得的结果进行了比较。日常可重复性研究和基于患者血清比较研究的数据显示,Stanbio和Medi-Chem钠试剂盒和Medi-Chem钾试剂盒在诊断上不能令人满意。 Stanbio钾试剂盒的结果是可重复的,但需要进一步改善,如患者血清分析中的非特异性所表明。 Coleman Jr.火焰光度计的不稳定性使得在这些测试中难以获得可接受的数据,特别是对于钠而言,而NIL火焰光度计具有良好的精度和准确性。