Automated Reaction-Rate Method for the Direct Determination of Total Serum Cholesterol by Use of the "CentrifiChem" Parallel Fast Analyzer

机译:使用“ CentrifiChem”平行快速分析仪直接测定总胆固醇的自动化反应速率法

获取原文
           

摘要

The direct procedure for determining serum cholesterol described by Wybenga et al. [ Clin. Chem. 16, 980 (1970)] can be used to assay 29 samples in 10-min reaction time by using the "CentrifiChem" parallel fast analyzer (Union Carbide). Commercial reference serum was used for standardization. Correlation of results with those obtained by the method of Abell et al. [ J. Biol. Chem. 195, 357 (1952)] for 54 samples yielded a regression line with a slope ( m ) of 0.897 and a y -intercept ( b ) of 231 mg/liter; the correlation coefficient ( r ) was 0.971. Compared with the manual procedure of Parekh and Jung [ Anal. Chem. 42, 1423 (1970)] for 146 samples, the corresponding values were m = 0.991, b = 27 mg/ liter, and r = 0.981. Compared to serum extracts assayed with the "AutoAnalyzer" (Standard Technicon Methodology N-24a), 186 samples gave corresponding values of m = 0.990, b = 92 mg/ liter, and r = 0.979. No bias significant at the 95% confidence level existed versus any of the three methods. Sera containing abnormally high bilirubin concentrations (50 mg/liter) or abnormally high triglyceride content (5,000 mg/liter) did not give significantly different cholesterol values on the CentrifiChem as compared with values obtained with either the AutoAnalyzer or the Parekh and Jung method.
机译:Wybenga等人描述的直接测定血清胆固醇的方法。 [临床。化学16,980(1970)]可通过使用“ CentrifiChem”平行快速分析仪(Union Carbide)在10分钟的反应时间内测定29个样品。商业参考血清用于标准化。结果与通过Abell等人的方法获得的结果的相关性。 [J. Biol。化学195,357(1952)]对54个样本产生的回归线的斜率(m)为0.897,y轴截距(b)为231 mg / L;相关系数(r)为0.971。与Parekh和Jung的手动程序比较[Anal。化学42,1423(1970)]对146个样品,相应的值是m = 0.991,b = 27 mg /升,r = 0.981。与使用“自动分析仪”(Standard Technicon Methodology N-24a)测定的血清提取物相比,有186个样品的相应值分别为m = 0.990,b = 92 mg /升和r = 0.979。与这三种方法中的任何一种相比,在95%置信度水平上都没有明显的偏差。与自动分析仪或Parekh and Jung方法获得的值相比,含有异常高的胆红素浓度(> 50 mg / L)或异常高的甘油三酸酯含量(> 5,000 mg / L)的血清在CentrifiChem上的胆固醇值没有显着不同。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号