首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >A Rapid Procedure for Catalase Determination in Blood and Tissue Samples with the Van Slyke Manometric Apparatus
【24h】

A Rapid Procedure for Catalase Determination in Blood and Tissue Samples with the Van Slyke Manometric Apparatus

机译:用Van Slyke测压仪快速测定血液和组织样品中的过氧化氢酶

获取原文
           

摘要

A precise and rapid gasometric procedure is described for catalase determination in blood and tissue samples. The enzyme activity assay is performed with the Van Slyke manometric apparatus. In the developed technicw, a 45-sec. reaction period is used, and a large liquid-gas interphase is provided with the purpose of preventing the oxygen diffusion from becoming a rate-limiting factor. Under the devised methodological conditions, the rate of oxygen evolution closely follows a parabolic curve which permits the calculation of the initial activity velocity of the enzyme. The oxygen evolvement is directly proportional to the quantity of catalase present in the analyzed sample. Less than 4 min. are required for each catalase measurement.
机译:描述了一种用于测定血液和组织样本中过氧化氢酶的精确而快速的胃压程序。酶活性测定是用Van Slyke压力计进行的。在发达的技术中,需要45秒。为了防止氧扩散成为限速因素,使用了反应时间,并设置大的气液界面。在设计的方法学条件下,氧的释放速率紧密遵循抛物线曲线,从而可以计算酶的初始活性速度。氧气逸出量与所分析样品中过氧化氢酶的量成正比。少于4分钟。每次过氧化氢酶测量都需要。