Effects of shielding the radiosensitive superficial organs of ORNL pediatric phantoms on dose reduction in computed tomography

机译:屏蔽ORNL小儿体模放射敏感性浅表器官对减少计算机断层扫描剂量的影响

获取原文
获取外文期刊封面目录资料

摘要

In computed tomography (CT), some superficial organs which have increased sensitivity to radiation, receive doses that are significant enough to be matter of concern. Therefore, in this study, the effects of using shields on the amount of dose reduction and image quality was investigated for pediatric imaging. Absorbed doses of breasts, eyes, thyroid and testes of a series of pediatric phantoms without and with different thickness of bismuth and lead were calculated by Monte Carlo simulation. Appropriate thicknesses of shields were chosen based on their weights, X-ray spectrum, and the amount of dose reduction. In addition, the effect of lead shield on image quality of a simple phantom was assessed quantitatively using region of interest (ROI) measurements. Considering the maximum reduction in absorbed doses and X-ray spectrum, using a lead shield with a maximum thickness of 0.4 mm would be appropriate for testes and thyroid and two other organs (which are exposed directly) should be protected with thinner shields. Moreover, the image quality assessment showed that lead was associated with significant increases in both noise and CT attenuation values, especially in the anterior of the phantom. Overall, the results suggested that shielding is a useful optimization tool in CT.Keywords: Image quality, lead and bismuth, Monte Carlo simulation, superficial organs' doses
机译:在计算机断层扫描(CT)中,一些对辐射敏感的浅表器官接受的剂量足够大,值得关注。因此,在这项研究中,对于小儿成像,研究了使用防护罩对剂量减少量和图像质量的影响。通过蒙特卡洛模拟法计算了一系列不带铋和铅的和不同厚度的小儿幻象的乳房,眼睛,甲状腺和睾丸的吸收剂量。根据其重量,X射线光谱和减少剂量的数量选择适当的防护层厚度。此外,使用关注区域(ROI)测量定量评估了铅屏蔽对简单体模图像质量的影响。考虑到吸收剂量和X射线光谱的最大减少,对于睾丸和甲状腺,使用最大厚度为0.4 mm的铅屏蔽层是合适的,甲状腺和其他两个器官(直接暴露)应使用较薄的屏蔽层进行保护。此外,图像质量评估显示铅与噪声和CT衰减值的显着增加有关,尤其是在幻影的前部。总体而言,结果表明屏蔽是CT中有用的优化工具。关键词:图像质量,铅和铋,蒙特卡洛模拟,浅表器官剂量

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号