首页> 外文期刊>Journal of Medical Physics/Association of Medical Physicists of India >Determination of absorbed dose to water for high-energy photon and electron beams-comparison of the standards DIN 6800-2 (1997), IAEA TRS 398 (2000) and DIN 6800-2 (2006)
【24h】

Determination of absorbed dose to water for high-energy photon and electron beams-comparison of the standards DIN 6800-2 (1997), IAEA TRS 398 (2000) and DIN 6800-2 (2006)

机译:测定高能光子和电子束在水中的吸收剂量-比较DIN 6800-2(1997),IAEA TRS 398(2000)和DIN 6800-2(2006)标准

获取原文
           

摘要

For the determination of the absorbed dose to water for high-energy photon and electron beams the IAEA code of practice TRS-398 (2000) is applied internationally. In Germany, the German dosimetry protocol DIN 6800-2 (1997) is used. Recently, the DIN standard has been revised and published as Draft National Standard DIN 6800-2 (2006). It has adopted widely the methodology and dosimetric data of the code of practice. This paper compares these three dosimetry protocols systematically and identifies similarities as well as differences. The investigation was done with 6 and 18 MV photon as well as 5 to 21 MeV electron beams. While only cylindrical chambers were used for photon beams, measurements of electron beams were performed using cylindrical as well as plane-parallel chambers. The discrepancies in the determination of absorbed dose to water between the three protocols were 0.4% for photon beams and 1.5% for electron beams. Comparative measurements showed a deviation of less than 0.5% between our measurements following protocol DIN 6800-2 (2006) and TLD inter-comparison procedure in an external audit.Keywords: Absorbed dose to water, DIN 6800-2 protocol (1997), DIN 6800-2 protocol (Draft National Standard 2006), external beam photons and electrons, IAEA TRS-398 protocol (2000)
机译:为了确定高能光子和电子束对水的吸收剂量,国际上采用了国际原子能机构(IAEA)行为准则TRS-398(2000)。在德国,使用德国剂量学协议DIN 6800-2(1997)。最近,DIN标准已被修订并发布为国家标准草案DIN 6800-2(2006)。它已广泛采用了实践准则的方法论和剂量学数据。本文系统地比较了这三种剂量学协议,并确定了相同点和不同点。使用6和18 MV光子以及5到21 MeV电子束进行了研究。虽然仅将圆柱形腔室用于光子束,但使用圆柱形以及平面平行的腔室进行电子束的测量。三种协议之间确定的水吸收剂量的差异对于光子束为0.4%,对于电子束为1.5%。比较测量结果表明,根据DIN 6800-2协议(2006)和外部审核中TLD的比对程序,我们的测量结果之间的偏差小于0.5%。关键词:水的吸收剂量,DIN 6800-2协议(1997),DIN 6800-2协议(国家标准草案2006),外部束光子和电子,IAEA TRS-398协议(2000)

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号