首页> 外文期刊>Journal of Medical Microbiology: An Official Journal of the Pathological Society of Great Britain and Ireland >Prospective evaluation of the VITEK MS for the routine identification of bacteria and yeast in the clinical microbiology laboratory: assessment of accuracy of identification and turnaround time
【24h】

Prospective evaluation of the VITEK MS for the routine identification of bacteria and yeast in the clinical microbiology laboratory: assessment of accuracy of identification and turnaround time

机译:对VITEK MS进行前瞻性评估以在临床微生物实验室中常规鉴定细菌和酵母菌:鉴定准确性和周转时间的评估

获取原文
       

摘要

This study assessed the accuracy of bacterial and yeast identification using the VITEK MS, and the time to reporting of isolates before and after its implementation in routine clinical practice. Three hundred and sixty-two isolates of bacteria and yeast, consisting of a variety of clinical isolates and American Type Culture Collection strains, were tested. Results were compared with reference identifications from the VITEK 2 system and with 16S rRNA sequence analysis. The VITEK MS provided an acceptable identification to species level for 283 (78 %) isolates. Considering organisms for which genus-level identification is acceptable for routine clinical care, 315 isolates (87 %) had an acceptable identification. Six isolates (2 %) were identified incorrectly, five of which were Shigella species. Finally, the time for reporting the identifications was decreased significantly after implementation of the VITEK MS for a total mean reduction in time of 10.52 h (P&0.0001). Overall, accuracy of the VITEK MS was comparable or superior to that from the VITEK 2. The findings were also comparable to other studies examining the accuracy of the VITEK MS, although differences exist, depending on the diversity of species represented as well as on the versions of the databases used. The VITEK MS can be incorporated effectively into routine use in a clinical microbiology laboratory and future expansion of the database should provide improved accuracy for the identification of micro-organisms.
机译:这项研究评估了使用VITEK MS进行细菌和酵母鉴定的准确性,以及在常规临床实践中实施分离株之前和之后报告分离株的时间。测试了362种细菌和酵母菌的分离株,其中包括各种临床分离株和美国典型培养物保藏中心菌株。将结果与来自VITEK 2系统的参考标识以及16S rRNA序列分析进行了比较。 VITEK MS为283(78%)分离株的物种水平提供了可接受的鉴定。考虑到常规临床护理可接受的属水平鉴定的生物,有315个分离株(87%)具有可接受的鉴定。错误地鉴定出6个分离物(占2%),其中5个是志贺氏菌。最后,实施VITEK MS后,报告鉴定的时间显着减少,总的平均时间减少了10.52小时(P <0.0001)。总体而言,VITEK MS的准确性与VITEK 2相当或更高。研究结果也与其他检查VITEK MS准确性的研究相当,尽管存在差异,这取决于所代表物种的多样性和种类。使用的数据库版本。 VITEK MS可以有效地并入临床微生物实验室的常规使用中,并且数据库的未来扩展将为鉴定微生物提供更高的准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号