A New Angle-Independent Doppler Ultrasonic Device for Assessment of Blood Flow Volume in the Extracranial Internal Carotid Artery

机译:一种新的独立于角度的多普勒超声仪,用于评估颅外颈内动脉血流量

获取原文

摘要

Objective. To evaluate a new angle-independent ultrasonic device for assessment of blood flow volume in the internal carotid artery. Methods. In vitro, a pulsatile pump was set to provide an outflow of physiological fluid at 500 mL/min through an 8-mm-diameter tube. Flow volume rates were measured 10 times by 10 different operators and compared with time-collected flow volume rates. In vivo, internal and common carotid artery blood flow volumes were measured in 28 volunteers by 2 operators using a FlowGuard device (Biosonix Ltd). Internal and common carotid artery diameters and blood flow volumes were also assessed by Duplex sonography and compared with FlowGuard measurements. In 10 volunteers, internal carotid artery blood flow volume changes in response to monitored breath manipulations were recorded. Results. In vitro, intraoperator variability was 4.04% (range, 2%–5.7%). The mean error rate ± SD was 3.54% ± 0.8% (range, 2.7%–5.2%). In vivo, the mean common carotid artery blood flow volume was 456 ± 39 mL/min (range, 417–583 mL/min) with a mean diameter of 6.7 ± 0.7 mm (range, 5.8–8.7 mm). The mean internal carotid artery blood flow volume was 277 ± 25 mL/min (range, 239–338 mL/min) with a mean diameter of 5 ± 0.5 mm (range, 4.1–6.1 mm). No significant difference was found between operators. Internal carotid artery diameter and blood flow volume measured by the FlowGuard were closely correlated with the results of Duplex sonography. Repeated shifts of end-tidal CO2 induced reproducible changes in internal carotid artery flow volume: 187.5 ± 18.1 mL/min at 26.8 ± 1.9 mm Hg and 382.1 ± 18.2 mL/min at 47 ± 2.2 mm Hg. Conclusions. The FlowGuard showed that volume flow studies in the internal carotid artery could be easily performed, with results compatible with those of previous clinical reports. Duplex comparative results and breath-induced changes in internal carotid artery flow volume justify further evaluation of the system.
机译:目的。为了评估一种新的与角度无关的超声设备,用于评估颈内动脉的血流量。方法。在体外,将搏动泵设置为以500 mL / min的速度通过直径8毫米的管流出生理液。流量率是由10位不同的操作员测量的10次,并与时间收集的流量率进行了比较。在体内,由2名操作人员使用FlowGuard设备(Biosonix Ltd)在28名志愿者中测量了颈内动脉和颈总动脉的血流量。还通过双重超声检查评估了颈内和颈总动脉的直径和血流量,并与FlowGuard测量结果进行了比较。在10名志愿者中,记录了响应于监测到的呼吸操作的颈内动脉血流量的变化。结果。在体外,术中变异性为4.04%(范围为2%–5.7%)。平均错误率±SD为3.54%±0.8%(范围为2.7%–5.2%)。在体内,平均颈总动脉血流量为456±39 mL / min(范围:417–583 mL / min),平均直径为6.7±0.7 mm(范围:5.8–8.7 mm)。颈内动脉平均血流量为277±25 mL / min(范围239–338 mL / min),平均直径为5±0.5 mm(范围4.1–6.1 mm)。运营商之间没有发现显着差异。 FlowGuard测量的颈内动脉直径和血流量与Duplex超声检查的结果密切相关。潮气末CO2的反复移动导致颈内动脉流量的可再现变化:在26.8±1.9 mm Hg时为187.5±18.1 mL / min,在47±2.2 mm Hg时为382.1±18.2 mL / min。结论。 FlowGuard显示可以轻松地进行颈内动脉的体积流量研究,其结果与以前的临床报告相符。双重比较结果和呼吸诱导的颈内动脉流量变化证明了对该系统进行进一步评估的合理性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号