Background With the availability of a wide variety of valve prostheses, surgeons are still searching for the ideal valve, striving for improved hemodynamics, durability and thrombogenicity. The purpose of this prospective, randomized, comparative study was to evaluate the early post-operative hemodynamic function and major clinical events in patients receiving CardiaMed? prosthetic valves in comparison to St. Jude Medical? (SJM) valves in the mitral valve position. Methods Between Sept 2013 and Sept 2015, 60 elective mitral valve replacement (MVR) patients in Ain Shams Hospital, Cairo, Egypt were divided into two groups of 30. Group I received the CardiaMed valve and Group II received the SJM valve. All patients were followed up at discharge from hospital, and at three and six months post-operatively. Results There were no statistical differences between groups regarding the demographic data, preoperative clinical and functional NYHA class, cardiac assessment or mitral valve pathology. By the end of the follow up, there was no valve-related co-morbidity. Early post-operative complications were seen in 4 patients (13.3%) in Group I and in 5 patients in Group II (16.7%), namely; re-exploration for bleeding, rhythm disturbance, or wound infection. The PPG and MPG were slightly higher in Group I (10.9?±?1.2 and 5.3?±?0.9) than in Group II (10.2?±?2.3 and 5.2?±?1.3); p = 0.798 and 0.107 respectively. There was no significant statistical difference between the groups regarding the post-operative echo follow-up data. Conclusions CardiaMed freely floating leaflet prostheses showed good hemodynamic characteristics. The prosthesis adequately corrects hemodynamics and is safe and no worse than the St. Jude Medical valve in the mitral valve position.
机译:背景技术随着各种各样的瓣膜假体的出现,外科医生仍在寻找理想的瓣膜,努力改善血液动力学,耐久性和血栓形成性。这项前瞻性,随机,比较研究的目的是评估接受CardiaMed?的患者的术后早期血流动力学功能和主要临床事件。人工瓣膜与St. Jude Medical相比? (SJM)阀位于二尖瓣位置。方法在2013年9月至2015年9月之间,将埃及开罗Ain Shams医院的60例二尖瓣置换术(MVR)患者分为两组,每组30例。第一组接受CardiaMed瓣膜,第二组接受SJM瓣膜。所有患者在出院时以及术后三个月和六个月都接受了随访。结果两组之间在人口统计学数据,术前临床和功能性NYHA分级,心脏评估或二尖瓣病变方面无统计学差异。在随访结束时,没有瓣膜相关的合并症。第一组中有4例患者(13.3%)出现了早期术后并发症,第二组中有5例患者(16.7%)出现了早期术后并发症。重新检查是否有出血,节律紊乱或伤口感染。 I组的PPG和MPG略高于II组(10.2±±2.3和5.2±±1.3)(10.9±1.2和5.3±0.9)。 p分别为0.798和0.107。关于术后回声随访数据,两组之间无统计学差异。结论CardiaMed自由浮动小叶假体具有良好的血液动力学特征。假体可以充分纠正血液动力学问题,并且安全且不会比二尖瓣位置的St. Jude Medical瓣膜差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号