首页> 外文期刊>Journal of the Formosan Medical Association =: Taiwan yi zhi >Impacts of 3D bone-to- implant contact and implant diameter on primary stability of dental implant
【24h】

Impacts of 3D bone-to- implant contact and implant diameter on primary stability of dental implant

机译:3D骨与种植体的接触和种植体直径对牙科种植体初级稳定性的影响

获取原文
           

摘要

Background/Purpose This study investigated the effects of three three-dimensional (3D) bone-to-implant contact (BIC) parameters—potential BIC area (pBICA), BIC area (BICA), and 3D BIC percentage (3D BIC%; defined as BICA divided by pBICA)—in relation to the implant diameter on primary implant stability, as well as their correlations were also evaluated. Methods Dental implants with diameters of 3.75, 4, 5, and 6?mm and artificial bone specimens were scanned by microcomputed tomography to construct 3D models for calculating pBICA, BICA, and 3D BIC%. Indexes of the primary implant stability including the insertion torque value (ITV), Periotest value (PTV), and implant stability quotient (ISQ) were measured after implants with various diameters were placed into bone specimens. The Kruskal–Wallis test, Wilcoxon rank-sum test with Bonferroni adjustment, and Spearman correlations were all performed as statistical and correlation analyses. Results The implant diameter significantly influenced pBICA and BICA, but not 3D BIC%. ITV and PTV were more sensitive to implant diameter than was ISQ. The coefficients of determination were high (0.92) for the correlations between pBICA (or BICA) and indexes of the primary implant stability. Conclusion This study revealed how the implant diameter and the three-dimensional (3D) BIC influence the primary stabilities of dental implant. ITV and PTV were more sensitively influenced by the implant diameter than ISQ. The pBICA and BICA seem to be more important than 3D BIC % for using wider implant in treatment plan, since those two parameters are highly predictive of variations in the primary stability of dental implant.
机译:背景/目的本研究调查了三个三维(3D)骨-植入物接触(BIC)参数的影响-潜在BIC面积(pBICA),BIC面积(BICA)和3D BIC百分比(3D BIC%;已定义(BICA除以pBICA)-与植入物直径有关的原始植入物稳定性以及它们之间的相关性也进行了评估。方法用微计算机断层扫描技术扫描直径为3.75、4、5和6?mm的牙科植入物以及人造骨标本,以构建3D模型以计算pBICA,BICA和3D BIC%。在将各种直径的植入物放入骨标本后,测量包括植入扭矩值(ITV),Periotest值(PTV)和植入物稳定性商(ISQ)在内的主要植入物稳定性指标。进行Kruskal–Wallis检验,带有Bonferroni调整的Wilcoxon秩和检验和Spearman相关性作为统计和相关性分析。结果植入物直径显着影响pBICA和BICA,但不影响3D BIC%。与ISQ相比,ITV和PTV对植入物直径更敏感。对于pBICA(或BICA)与初始植入物稳定性指标之间的相关性,确定系数较高(> 0.92)。结论这项研究揭示了种植体直径和三维(3D)BIC如何影响牙齿种植体的主要稳定性。与ISQ相比,ITV和PTV受植入物直径的影响更为敏感。对于在治疗计划中使用更大的种植体,pBICA和BICA似乎比3D BIC%更重要,因为这两个参数可以高度预测牙齿种植体的基本稳定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号