首页> 外文期刊>Journal of the Brazilian Chemical Society >Strategies to improve accuracy and sensitivity in phosphorus determinations by inductively coupled plasma quadrupole mass spectrometry
【24h】

Strategies to improve accuracy and sensitivity in phosphorus determinations by inductively coupled plasma quadrupole mass spectrometry

机译:通过电感耦合等离子体四极杆质谱法提高磷测定的准确性和灵敏度的策略

获取原文
           

摘要

The collision/reaction interface with He introduced through the skimmer at 100 mL min-1 is employed to improve accuracy in 31P+ determinations by ICP-QMS. The 31P16O+ determination in cool plasma conditions result in limits of detection 55-fold lower. This strategy may be an interesting alternative for high ionization potential elements such as P.
机译:通过撇油器在100 mL min-1下与He引入的碰撞/反应界面可提高ICP-MS进行31P +测定的准确性。在冷血浆条件下的31P16O +测定导致检测极限降低了55倍。对于高电离势元素(例如P),此策略可能是有趣的替代方法。