首页> 外文期刊>Journal of radiation research >The dosimetric impact of respiratory breast movement and daily setup error on tangential whole breast irradiation using conventional wedge, field-in-field and irregular surface compensator techniques
【24h】

The dosimetric impact of respiratory breast movement and daily setup error on tangential whole breast irradiation using conventional wedge, field-in-field and irregular surface compensator techniques

机译:使用常规楔形,场内和不规则表面补偿器技术,呼吸乳房运动和每日设置误差对切向全乳辐射的剂量学影响

获取原文
       

摘要

A total of 16 consecutive patients with early-stage breast cancer who received WBI were chosen for this study. Right breasts were treated in 8 patients and left breasts were treated in the other 8. For all patients, the prescribed dose was 50 Gy in 25 fractions. CT image sets were transferred to the Eclipse treatment planning system (Varian Medical Systems, Palo Alto, CA, USA) for contouring and treatment planning. The patient outline, clinical target volume (CTV), planning target volume (PTV) and whole lung were contoured on each CT slice. The CTV was defined as the entire breast tissue including glandular breast and surrounding soft tissue. The PTV was defined as the CTV plus an expansion of 5 mm in all directions except for the external skin surface. The lung volume was automatically generated using the auto-contouring tool of the treatment planning system. The evaluated CTV (CTVevl) was defined as the volume of the CTV enclosed by contours drawn 5 mm below the skin surface to eliminate the region of dose build-up [6]. We performed treatment planning using two tangential fields for breast irradiation with three different techniques: CW, FIF and ISC. For field-in-field (FIF), we followed the procedures reported by Kestin et al. [3]. The monitor unit (MU) weights of the open radiation field and those of the sub-fields were adjusted to remove hot spots in each plan. Taking into account dose output stability, we selected sub-fields with more than 6 MU. Irregular surface compensator (ISC), a type of electronic compensator, is a feature installed in Eclipse which enables improved dose homogeneity for irregular surface shapes. The X-ray fluence distribution required to produce an isodose surface perpendicular to the central axis at a specified depth is calculated by Eclipse and then delivered using a dynamic multileaf collimator (MLC). If the dose is not sufficiently homogenous, painting the fluence map can modify fluence distribution to achieve better dose homogeneity. The region enclosed by the 107 % isodose contours is erased by manual fluence painting. The fluence value of the superficial region is also applied in the flash region in order to ease MLC movement in the flash region. A dose–volume histogram (DVH) analysis was performed for all regions of interest. The following dose indices were used to evaluate the plan quality where relative dose means the ratio of the received to the prescribed dose, and relative volume shows the fraction of the whole contoured volume of each region of interest. Based on the CT image set of each patient, an ‘original plan' was created using the usual clinical planning technique of our Institution. Our approach for the evaluation of dose variation was to shift the isocenter of each original plan in every direction and estimate the variation of each dose index with isocenter shift. To analyze movement in the AP direction, the variation of dose indices caused by respiratory motion and setup error were calculated for a single fraction using the method described below. Assuming that daily setup error has a normal distribution for systematic and random error, the dose indices after all fractions () are calculated using , as follows; (3) where is the normal distribution function of the systematic setup error E. M indicates the overall mean error and Σ is the standard deviation (SD) of systematic setup error. The normal distribution function of the random setup error e is defined as where σ is the SD of the random setup error [17]. All patient data were used in the statistical calculation. Student's t-test or Welch's t-test (two-tailed) was applied to compare each dose index between irradiation techniques and to investigate the effects of respiratory motion and setup error (i.e. comparing the dose indices on original and isocenter-shift plans, and ). A P value of ≤0.05 was considered statistically significant. Table 1 summarizes the mean value and 1 SD of each dose index obtained from the original plan. Table 2 shows the dose indices for CTVevl, lung and body with isocenter shift in the lateral and SI directions. For all techniques, isocenter shifts of 0.5 cm along these axes showed no significant variation from the original plans for most of the indices. However, for the lateral shift in ISC, which indicates a lateral shift away from the midline, V95 significantly decreased from 99.2% to 96.9% and D95 also decreased from 97.3% to 96.0%. The reason is that if a patient shifts away from the midline, the target moves into the flash region (see Fig. 1), where the fluence intensity is small to ensure a smooth movement of MLC, and not sufficient to give the prescribed dose to the target. Table 4 shows the dose indices, taking respiratory motion into account for a single fraction treatment (). Because we confirmed that the dose indices were unaffected by isocenter shifts in the lateral and SI directions (Table 2), both respiratory motion and s
机译:总共选择了连续16例接受WBI治疗的早期乳腺癌患者。右乳房8例,其他8例。所有患者的处方剂量为50 Gy(25份)。 CT图像集已传输到Eclipse治疗计划系统(Varian Medical Systems,美国加利福尼亚州帕洛阿尔托)进行轮廓和治疗计划。在每个CT切片上绘制患者轮廓,临床目标体积(CTV),计划目标体积(PTV)和整个肺部轮廓。 CTV被定义为整个乳腺组织,包括腺性乳腺和周围的软组织。 PTV定义为CTV加上皮肤外表面在所有方向上的5毫米膨胀。使用治疗计划系统的自动轮廓工具自动生成肺部容积。评估的CTV(CTV evl )定义为CTV的体积,该体积被皮肤表面以下5 mm绘制的轮廓所包围,以消除剂量累积区域[6]。我们使用两个切线场对乳房进行放射治疗,并采用三种不同的技术进行治疗计划:CW,FIF和ISC。对于现场(FIF),我们遵循Kestin等人报告的步骤。 [3]。调整开放辐射场和子场的监视单元(MU)权重,以消除每个计划中的热点。考虑到剂量输出的稳定性,我们选择了6 MU以上的子场。不规则表面补偿器(ISC)是一种电子补偿器,是Eclipse中安装的一项功能,可以提高不规则表面形状的剂量均匀性。通过Eclipse计算在指定深度处产生与中心轴垂直的等剂量表面所需的X射线通量分布,然后使用动态多叶准直仪(MLC)进行传递。如果剂量不够均匀,则绘制注量图可以修改注量分布以实现更好的剂量均匀性。 107%等剂量线轮廓所包围的区域通过手动注量涂抹消除。为了减轻MLC在闪光区域中的移动,还在闪光区域中应用了表面区域的注量值。对所有感兴趣区域进行剂量-体积直方图(DVH)分析。以下剂量指数用于评估计划质量,其中相对剂量表示所接收剂量与处方剂量之比,相对体积表示每个感兴趣区域的总轮廓体积的分数。根据每位患者的CT影像集,使用我们机构常用的临床计划技术创建了一个“原始计划”。我们评估剂量变化的方法是在每个方向上移动每个原始计划的等中心点,并估计每个剂量指数随等中心点变化的变化。为了分析在AP方向上的运动,使用下述方法针对单个分数计算了由呼吸运动和设置误差引起的剂量指数变化。假设每日设置误差具有系统误差和随机误差的正态分布,则使用来计算所有分数()之后的剂量指数,如下所示; (3)其中是系统设置误差E的正态分布函数。M表示总体平均误差,Σ是系统设置误差的标准偏差(SD)。随机设置误差e的正态分布函数定义为,其中σ是随机设置误差的SD [17]。所有患者数据均用于统计计算。应用学生t检验或韦尔奇t检验(两尾)比较辐射技术之间的每个剂量指数,并研究呼吸运动和设置误差的影响(即比较原始和等中心线偏移计划上的剂量指数,以及)。 P值≤0.05被认为具有统计学意义。表1总结了从原始计划获得的每个剂量指数的平均值和1 SD。表2列出了CTV evl ,肺和身体在横向和SI方向等中心点偏移的剂量指数。对于所有技术,大多数指标沿这些轴的0.5 cm等中心线位移均与原始计划无明显差异。但是,对于ISC中的横向偏移(表示偏离中线),V 95 从99.2%显着降低到96.9%,D 95 也从97.3%降低。 %至96.0%。原因是,如果患者偏离中线,则目标会进入闪光区域(参见图1),该区域的注量强度很小,无法确保MLC的平稳运动,并且不足以向患者提供指定剂量目标。表4显示了剂量指数,其中考虑了单次治疗的呼吸运动()。因为我们确认剂量指数不受横向和SI方向上等中心线位移的影响(表2),所以呼吸运动和s

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号