Palpebral Fissure Length Measurement: Accuracy of the FAS Facial Photographic Analysis Software and Inaccuracy of the Ruler

机译:睑裂长度测量:FAS面部图像分析软件的准确性和尺子的不准确性

获取原文
           

摘要

BackgroundAccurate fetal alcohol spectrum disorder diagnoses require accurate facial measurement. The Fetal Alcohol Syndrome (FAS) Facial Photographic Analysis Software was developed to overcome measurement error known to occur with ruler measurement of the PFL. Recent publications have queried the Software’s accuracy.Objectives1)Demonstrate the Software’s ability to accurately measure a PFL from a 2-dimensional digital facialphotograph. 2) Demonstrate the frequency and magnitude of error when the PFL is measured directly by clinicians using a ruler.MethodsObjective 1: PFLs of mannequins were measured using the Software and a sliding digital caliper, with the latter serving as the gold-standard accurate measure. Mannequins allowed the caliper prongs to be placed directly on the landmarks that define the PFL. Objective 2: PFLs of 1,027 patients evaluated at the University of Washington FAS Diagnostic & Prevention Network were measured with the Software and directly by one or two clinicians using a ruler.ResultsObjective 1: The Software derived PFLs that were identical to or within 0.2 mm of the caliper measures. Objective 2: There was tremendous inter-rater variability in PFLs measured by clinicians using a hand held ruler. Seventy-seven percent of patients had their PFLs measured incorrectly (greater than 1 mm error) by at least one of the two clinicians using a ruler.ConclusionThe FAS Facial Photographic Analysis Software measures the PFL with the same accuracy as a sliding digital caliper, as it was programmed to do. Direct measurement of the PFL with a ruler is very prone to error.
机译:背景技术准确的胎儿酒精频谱障碍诊断需要准确的面部测量。开发了胎儿酒精综合症(FAS)面部图像分析软件,以克服已知的PFL标尺测量中出现的测量误差。最近的出版物质疑该软件的准确性。目标1)演示该软件从二维数码面部照片准确测量PFL的能力。 2)演示临床医生使用尺子直接测量PFL时的错误频率和幅度方法方法目标1:使用软件和滑动数字卡尺测量人体模型的PFL,后者用作金标准精确测量。人体模型允许将卡钳插脚直接放置在定义PFL的地标上。目标2:使用该软件并直接由一名或两名临床医生使用尺测量在华盛顿大学FAS诊断与预防网络评估的1,027例患者的PFL结果结果目的1:该软件衍生的PFL等于或小于0.2 mm卡尺的措施。目标2:临床医生使用手持尺测量的PFL存在很大的评估者间差异。两名临床医生中至少有一位使用尺子对77%的患者的PFL进行了错误的测量(误差大于1 mm)。结论FAS面部图像分析软件测量PFL的精度与滑动数字卡尺相同,它被编程做。用直尺直接测量PFL非常容易出错。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号