Neutron flux distributions in the Kyoto University Critical Assembly were measured with a position-sensitive 3He proportional counter. The authors examined the results of the first attempt to apply the position-sensitive counter for such measurement in the critical assembly. Some problems to be solved were drawn from the present examination for the further development of this type of counters. The counter had a sensitive length of 118 cm and an outer diameter of 2.5 cm. The counting gas was a mixture of 3He (270 Pa) and CF4 (3 x 104 Pa). The observed distribution was very similar to that measured with the conventional gold wire activation method. The position resolution was 2.7 cm in the experimental condition. The measurable neutron flux (at peak position) was limited to the order of 104 n?cm-2?s-1 due to the pileup of signal pulses. The counter was a useful tool to quickly measure the flux distribution in the assembly.