首页> 外文期刊>Journal of Optometry >Anterior vaulting of single piece in-the-bag intraocular lens after blunt trauma
【24h】

Anterior vaulting of single piece in-the-bag intraocular lens after blunt trauma

机译:钝性创伤后单件袋装人工晶状体的前拱

获取原文
           

摘要

A 64-year-old man complained of difficulty with far vision in the right eye after blunt trauma by his hand during sleep. He had undergone uneventful phacoemulsification and intraocular lens (IOL) implantation of both eyes one week before the event. On slit lamp examination, the anterior chamber was shallow compared to postoperative one day or the left eye. After mydriasis, IOL position was perfect, within the capsulorrhexis margin, and no haptics were protruded. SE of his right eye was -4.00 diopter after trauma, although spherical equivalent (SE) of both eyes was near emmetropia one day after cataract surgery. We decided on IOL reposition or exchange. Intraoperatively, we pushed the IOL posteriorly; the IOL was vaulted posteriorly, as expected. After surgery, his SE was +0.25 with the same diopter IOL. We report the case on the possibility of 4-point haptic IOL vaulting anteriorly by blunt trauma within the capsular bag.
机译:一名64岁的男子抱怨在睡觉时手被钝伤后右眼远视困难。在事件发生前一周,他接受了双眼的超声乳化和人工晶状体(IOL)植入。在裂隙灯检查中,与术后一天或左眼相比,前房较浅。散瞳后,IOL位置完美,在闭经边缘,并且没有突。尽管白内障手术后一天,两只眼睛的球当量(SE)接近正视,但右眼的SE在受伤后为-4.00屈光度。我们决定了人工晶体的重新定位或交换。术中,我们向后推人工晶体;如预期的那样,IOL在后部进行了拱顶。手术后,使用相同的屈光度IOL,他的SE为+0.25。我们报道了由囊袋内钝性损伤导致四点触觉人工晶状体向前拱形的可能性。