首页> 外文期刊>Journal of Diabetes Science and Technology >Continuous Glucose Monitoring: Real-Time Algorithms for Calibration, Filtering, and Alarms
【24h】

Continuous Glucose Monitoring: Real-Time Algorithms for Calibration, Filtering, and Alarms

机译:连续葡萄糖监测:用于校准,过滤和警报的实时算法

获取原文

摘要

Algorithms for real-time use in continuous glucose monitors are reviewed, including calibration, filtering of noisy signals, glucose predictions for hypoglycemic and hyperglycemic alarms, compensation for capillary blood glucose to sensor time lags, and fault detection for sensor degradation and dropouts. A tutorial on Kalman filtering for real-time estimation, prediction, and lag compensation is presented and demonstrated via simulation examples. A limited number of fault detection methods for signal degradation and dropout have been published, making that an important area for future work.
机译:审查了在连续血糖监测仪中实时使用的算法,包括校准,噪声信号过滤,降血糖和高血糖警报的葡萄糖预测,补偿毛细血管血糖到传感器时滞以及传感器故障和掉线的故障检测。通过仿真示例介绍并演示了用于实时估计,预测和滞后补偿的卡尔曼滤波教程。已经发布了数量有限的用于信号降级和丢失的故障检测方法,这使其成为未来工作的重要领域。