Simulation of impedance measurements at human forearm within 1 kHz to 2 MHz

机译:在1 kHz至2 MHz范围内模拟人类前臂的阻抗测量

获取原文
       

摘要

This work presents a simulation analysis of the bioimpedance measurements at human forearm. The Ansys? High Frequency Structure Simulator (HFSS) has been used to analyze the electrical response of a section of human forearm with three domains of di-electric behavior- fat, muscle and artery (blood). The impedance values were calculated as the ratio of the output voltage at the electrodes to the applied known current (1mA). A model was developed and was simulated for impedance values obtained within a frequency range of 1 kHz to 2MHz. The measurements were done at three instances of radial artery diameter. The maximum resistance and reactance values were calculated as 445Ω and 178.5Ω, 356Ω and 138Ω, and 368Ω and 144.3Ω for diameters 2.3mm, 2.35mm, and 2.4mm respectively. The set of impedance values obtained followed Cole-plot trend. The results obtained were found to be in excellent agreement with the Cole theory. The set of values obtained at three different diameters reflected the effect of blood flow on impedance values.
机译:这项工作提出了对人类前臂生物阻抗测量值的模拟分析。 Ansys?高频结构模拟器(HFSS)已用于分析人类前臂部分具有三个介电行为域(脂肪,肌肉和动脉(血液))的电响应。将阻抗值计算为电极上的输出电压与施加的已知电流(1mA)之比。开发了一个模型,并针对在1 kHz至2MHz频率范围内获得的阻抗值进行了仿真。在at动脉直径的三种情况下进行测量。对于直径为2.3mm,2.35mm和2.4mm的最大电阻和电抗值分别计算为445Ω和178.5Ω,356Ω和138Ω,以及368Ω和144.3Ω。获得的一组阻抗值遵循科尔图趋势。发现获得的结果与科尔理论非常吻合。在三个不同直径处获得的一组值反映了血流对阻抗值的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号