Background: The study evaluated the impact of interferences on the analytical specificity of three commercial and commonly used creatinine methods (two Jaffe and one enzymatic).Methods: Manufacturer creatinine methods plus modified methods were tested with the following interferences: spiking serum with bilirubin, albumin, glucose, hemoglobin and lipid, and patient sera with maximum concentrations of bilirubin, 1,090 μmol/l and protein, 117.8 g/l.Results: Hemoglobin, 7.5 g/l and lipaemic with triglyceride concentration of 6.27 mmol/l, did not interfere with all assays. Glucose 33.3 mmol/l increased creatinine recovery for Dimension method. Samples spiked with bilirubin imparted a negative bias for Dimension and Architect methods but imparted a positive bias for Vitros assay. However, using patient sera, negative bias with bilirubin was found for all methods, from which Architect method gave the highest effect (RSUP2/SUP=0.861), followed by Vitros (RSUP2/SUP=0.239) and Dimension (RSUP2/SUP=0.163). Protein provided the positive bias for all creatinine measurements that increased with increasing concentration (RSUP2/SUP ranging from 0.104 to 0.182, P0.0001). Addition of sodium dodecyl sulfate (SDS) in alkaline-picrate reagent reduced the effect of bilirubin and protein for kinetic Jaffe method. Although adding potassium ferricyanide was well effective for eliminating negative interference of bilirubin, it was prone to interference from protein.Conclusions: Endogenous interferences continue to plague creatinine accuracy measurement in both Jaffe and enzymatic methods, and consequentially the estimated glomerular filtration rate. The addition of SDS to the alkaline-pirate reagent was shown to be effective in reducing bilirubin and protein interferences. J. Clin. Lab. Anal. 24:123-133, 2010. ? 2009 Wiley-Liss, Inc.
机译:背景:本研究评估了干扰对三种商业和常用肌酐方法(两种Jaffe和一种酶促方法)的分析特异性的影响。方法:对制造商的肌酐方法和改良方法进行了以下干扰测试:血清掺有胆红素,白蛋白,葡萄糖,血红蛋白和脂质以及患者血清中的最大胆红素浓度为1,090μmol/ l,蛋白质为117.8 g / l。结果:血红蛋白7.5 g / l和血脂性甘油三酸酯浓度为6.27 mmol / l,无干扰与所有检测。 Dimension方法中,葡萄糖> 33.3 mmol / l增加了肌酐的回收率。掺有胆红素的样品对Dimension和Architect方法产生负偏倚,但对Vitros分析产生正偏倚。然而,使用患者血清,所有方法均发现胆红素呈负偏倚,其中Architect方法效果最高(R 2 = 0.861),其次是Vitros(R 2 = 0.239)和尺寸(R 2 = 0.163)。蛋白质为所有肌酐测量提供了正偏差,并随浓度的增加而增加(R 2 的范围从0.104到0.182,P <0.0001)。在碱性苦味剂中添加十二烷基硫酸钠(SDS)会降低动力学Jaffe方法的胆红素和蛋白质的作用。尽管加入铁氰化钾可以很好地消除胆红素的负面干扰,但它容易受到蛋白质的干扰。结论:内源性干扰继续困扰着Jaffe和酶法的肌酐准确性测量,因此估计了肾小球滤过率。已证明将SDS添加到碱性海盗试剂中可有效减少胆红素和蛋白质干扰。 J.临床。实验室肛门24:123-133,2010。 2009 Wiley-Liss,Inc.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号